5-5
Alaris
®
SpO
2
Modules
Technical Service Manual
CORRECTIVE MAINTENANCE
5.2.3 Removing
Frame
Assembly
1. Remove screws (
6
) attaching Frame
Assembly to Front Case Assembly.
2. Remove
IUI
Bracket Strap screw.
3. Separate Frame Assembly from Front
Case Assembly.
During Reassembly:
Align Frame and Front Case Assemblies as
follows:
• Engage top flange of Frame with square
slot on Status Indicator Lens.
• Align and engage display connectors
(
2
) with Power Supply Board and Logic
Board connectors.
• Ensure Patient Board connector is
centered within square cutout on Front
Case.
• Position Keypad Ground Wire between
IUI
Bracket Strap and screw.
5.2
DISASSEMBLY / REASSEMBLY
(Continued)
Frame Assembly
Front Case Assembly
Logic Board
IUI Bracket Strap Screw
IUI Bracket Strap
Keypad Ground Wire
(positioned between IUI Bracket Strap and screw)