4

CALIBRATION PROCEDURE Continued

 

Place test weight equal to ten percent of the scales capacity over each of the four load cells, one at 
a time.  Record each displayed weight.  If difference in displayed weight exceeds ±.05 percent of 
the test weight, trimming adjustments for your load cells must be performed.  Refer to TRIM 
ADJUSTMENT Section for further instruction.  If all four weight readings are within the allowable 
tolerance (±.05 percent of test weight), weight indicator should be re-calibrated in accordance with 
the indicator manual.  
 
 

TRIM ADJUSTMENT 

The purpose of adjusting trim is so the same weight reading will be displayed regardless of where 
the load is placed on the scale deck.  All scales are corner trimmed before leaving the factory.  In 
the event of re-calibration or load cell replacement, follow instructions below to re-trim corners.  It is 
assumed the other three load cells are already corner sealed to agree with each other.  If not, 
please refer to next page for further instruction. 
 
1.  Remove access plate located on scale deck. 
2.  Remove junction box access cover. 
3.  Refer to Figure No. 3 for load cell location reference. 
4.  Turn trim potentiometer for the load cell needing re-

calibration clockwise 25 turns, or until a click is heard 
indicating the end of the potentiometer adjustment. 

5.  Place a test weight on each corner, one at a time, 

and record the displayed weight for each corner. 

6.  Place test weight on the corner with the highest 

reading and adjust the appropriate trim 
potentiometer until the weight reading agrees with 
the lowest reading obtained in Step 5. 

7.  Repeat Step 6 until all corners have the same weight 

reading. 

8.  Replace screws securing the junction box cover and 

access plate. 

           

 

 

 

 

 

Figure No. 3

 

 
If two or more load cells must be replaced or scale needs an over-all resealing of each corner, 
these procedures should be followed. 
 
1.  Remove access plate located on scale deck.  Remove junction box access cover. 
2.  Turn all potentiometers clockwise 25 turns or until a click is heard.  Now turn all potentiometers 

counterclockwise 1 (one) turn. 

3.  Place test weight of at least 10 percent of scales capacity on each corner, one at a time, and 

record each displayed weight. 

4.  Place test weight on corner with the highest displayed weight and adjust appropriate trim 

potentiometer counter-clockwise until corner reading agrees with lowest weight reading. 

5.  Repeat Step No. 4 until all corners are equal. 
6.  Re-calibrate weight indicator following procedures in indicator manual. 
7.  Replace screws securing the junction box cover and access plate. 
 

EXAMPLE: 

 CORNER 

NO. 

1000 

lb 

TEST 

WEIGHT 

1 2 3 4 

DISPLAYED 

WEIGHT 

READING  1020 1016 1017 1021 

 
1.  Place test weight on each corner, one at a time, and record each displayed weight. 
2.  Place test weight on corner no. 1 and adjust no. 1 trim potentiometer to reduce displayed 

weight to the lowest corner reading (No. 2 in example). 

3.  Repeat step 2 with corners no. 3 and 4 until all corners are in agreement. 
4.  Calibrate weight indicator following instructions provided in its manual. 

Summary of Contents for Floor Hugger

Page 1: ...c Platform Scale Owner s Manual CARDINAL SCALE MFG CO 1905 M142 O1 Rev D 203 E Daugherty Webb City MO 64870 USA 05 01 Ph 417 673 4631 Fax 417 673 2153 Printed in USA www cardinalscale com Technical Support Ph 866 254 8261 tech cardet com ...

Page 2: ...2 ...

Page 3: ...ability is assumed with respect to the use of the information contained herein TABLE OF CONTENTS Introduction 1 Installation 2 Calibration Procedure 3 Trim Adjustment 4 Trouble Shooting Guide 5 Interconnection Diagram 6 FCC COMPLIANCE STATEMENT This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions in this manual may c...

Page 4: ...al Scale should remain packed until you are ready to install it protecting it while being stored Remove shipping bands and packing material Make certain all packing material is removed Remove wooden box containing load cell cable from pallet Place on scale platform until further instructions are given BE CERTAIN scale location is level and free of debris Using the lifting eye bolt provided move sc...

Page 5: ...able to weight indicating instrument following instructions provided in instrument manual Make certain cable is not placed near heating or cooling ducts relay panels ovens or other types of electrical equipment All cables should be routed out of the way of normal traffic and secured to prevent accidental damage to operator and or instruments After connection has been made power up and zero weight ...

Page 6: ...t weight on the corner with the highest reading and adjust the appropriate trim potentiometer until the weight reading agrees with the lowest reading obtained in Step 5 7 Repeat Step 6 until all corners have the same weight reading 8 Replace screws securing the junction box cover and access plate Figure No 3 If two or more load cells must be replaced or scale needs an over all resealing of each co...

Page 7: ... SOLUTION Scale reads accurate weight to a point but will not register beyond that point Obstruction between floor and bridge Remove obstruction Cannot obtain repeatable readings A Scale not level B Bad or improper wiring C Bad load cell D Bad trim circuit board A Level Scale B Check wiring solder connections C Replace load cell D Replace trim circuit board One corner will not register correct wei...

Page 8: ...6 INTERCONNECTION DIAGRAM Figure No 4 ...

Page 9: ...7 ...

Page 10: ...8 ...

Reviews: