SETUP AND CALIBRATION, CONT.

UNLOAD WEIGHT

With the display showing 

uLd

uLd

, remove the 2 lb test weight from the scale tray.  Allow 

approximately five seconds for the scale to stabilize, then press the 

ON/TARE

key to continue 

the calibration.  Dashes will be displayed from the left to the right of the display momentarily, 
then the display will change to 

unt

unt

.

If

ErA

ErA

 is displayed during the 

LOd

LOd

 or 

uld

uld

 calibration sequence, remove the test weight 

from the scale and verify that it is the correct weight (2 lbs) or insure that the scale is 
empty and nothing is touching the scale tray.  The display will return to 

LOd

LOd

 or 

uld

uld

allowing you to continue with calibration and setup. 

WEIGHING UNITS

With the display showing 

unt

unt

, press the 

WT/UNIT

 key to display the current setting for the 

default weighing unit.  The display will show 

-- -- --

 and turn on the red indicator for the weight 

unit (oz or gm) or 

1

1

 and the red indicator for (gm) for the "Grams only" mode.  Press the 

WT/UNIT

 key until the desired weight unit is indicated.  Press the 

ON/TARE

key to save the 

setting.

AUTOMATIC SHUTOFF

The display will show 

ASH

ASH

.  Press the 

WT/UNIT

 key to display the current setting.  Press the 

WT/UNIT

 key until the desired number of minutes of inactivity before the scale will 

automatically shutoff is displayed.  Values of 1, 2, 5, 10, or 15 minutes or OFF are available. 
Selecting OFF will disable the automatic shutoff feature. Press the 

ON / TARE 

key to save the 

setting.

POWER UP ZERO 

The display will show 

PuO

PuO

.  Press the 

WT/UNIT

 key to display the current setting.  Press the 

WT/UNIT

 key to select 

YES

YES

to enable or

 no 

 no 

to disable.  Press the 

ON / TARE 

key to save the 

setting.

DONE

The scale will briefly display 

dnE

dnE,

then reset to the Operation weight mode.

SETUP REVIEW

The Setup Review mode allows you to review and change several operational parameters as 
desired without having to enter the setup and calibration mode.  To enter the setup review mode, 
you need to turn the scale off by pressing the 

WT/UNIT

 and 

ON-TARE

 keys simultaneously and

then pressing and holding the 

ON-TARE

key until the red indicator of the selected weight units 

begins to flash.  After the red indicator begins to flash, release the 

ON-TARE

 key.  The display will 

show

unt

unt

, the prompt for the weighing units.  Refer to the instructions listed in the Setup and 

Calibration section  for information on how to change these parameters.  The parameters in the 
setup review that can be change are as follows:

WEIGHING UNITS

AUTOMATIC SHUTOFF

Location of
Calibration
Switch

Summary of Contents for Detecto PS-5

Page 1: ... DIGITAL PORTION SCALE PS 5 OWNER S Manual CARDINAL SCALE MFG CO 8528 M013 O1 Rev G PO BOX 151 WEBB CITY MO 64870 Printed in USA 09 00 PH 417 673 4631 FAX 417 673 5001 Web Site http www detectoscale com ...

Page 2: ...2 ...

Page 3: ...vidual accuracy skill and caution For this reason the Seller is not able to guarantee the result of any procedure contained herein Nor can they assume responsibility for any damage to property or injury to persons occasioned from the procedures Persons engaging the procedures do so entirely at their own risk TABLE OF CONTENTS Specifications 1 Installation 2 Battery Replacement 2 Operation 2 Automa...

Page 4: ...d batteries and replace with fresh ones Make certain the and ends of the batteries match the positions shown on the scale circuit boards as shown Replace the battery compartment cover and retaining screw OPERATION 1 TURN THE SCALE ON Press the ON TARE key 2 SELECT THE WEIGHING UNIT Press the WT UNIT key to alternate between ounces and gram weight displays A red indicator light will illuminate to s...

Page 5: ...h the drawing on the circuit board If using the optional AC power adapter insert the plug into the jack on the back of the scale BEGIN SETUP and CALIBRATION 1 Making certain the scale tray is empty press and hold the setup and calibration switch located on the lower right side of the scale Refer to the figure on the next page 2 Press the ON TARE key to turn the scale on Release the setup and calib...

Page 6: ... setting Press the WT UNIT key until the desired number of minutes of inactivity before the scale will automatically shutoff is displayed Values of 1 2 5 10 or 15 minutes or OFF are available Selecting OFF will disable the automatic shutoff feature Press the ON TARE key to save the setting POWER UP ZERO The display will show PuO PuO Press the WT UNIT key to display the current setting Press the WT...

Page 7: ...2 ...