background image

3

DISPLAY CODES

CALIBRATION REQUIRED -

 Indicates improper stored calibration data, calibration 

is necessary.

OVER CAPACITY -

 Indicates a weight exceeding the capacity of scale has  been 

placed on the scale tray.

ERROR -

 NOV-RAM test failure.

ANALOG FAILURE -

 This could mean a damaged or defective load cell, a failed 

component in the analog circuity of the scale or the platform isn't on or is on 
incorrectly.

SETUP AND CALIBRATION

Your scale has been thoroughly tested and calibrated before being shipped to you and should not 
require adjustment.  In the event the scale should need re-calibration, the following calibration 
procedure should be performed.  To maintain the scale's high degree of accuracy, it is 
recommended that a qualified technician with certified test weights re-calibrate the scale.

In order to perform Setup and Calibration, it is necessary to remove the "Do Not Remove" sticker 
from the lower right side of the scale.  With the sticker removed, locate the Setup and Calibration 
switch in the opening closest to the front of the scale. Refer to the figure on the next page.

NOTE: If operating from batteries, install the batteries making certain that they are 
positioned in accordance with the drawing on the circuit board.  If using the 
optional AC power adapter, insert the plug into the jack on the back of the scale. 

BEGIN SETUP and CALIBRATION

1. Making certain the scale tray is empty, press and hold the setup and calibration switch located 
on the lower right side of the scale.  Refer to the figure on the next page. 

2. Press the

ON/TARE

key to turn the scale on.  Release the setup and calibration switch when 

the display is showing 

SCA

SCA.

SCALE MODEL

The display will show 

SCA

SCA

.  Press the 

WT/UNIT

 key to display the current setting.  For a PS-5

the display should show 

P

P--5

5

.  Press the

WT/UNIT

 key until 

P

P--5

5

 is displayed.  Press the 

ON/TARE

key to save the setting.

CALIBRATION

The display will show 

CAL

CAL

.  Press the 

WT/UNIT

 key to display the current setting.  To calibrate 

the scale, press the 

WT/UNIT

 key until 

YES

YES

is displayed, then press the 

ON/TARE

key.  The 

display will change to 

LOd

LOd

.

To skip calibration, press the 

WT/UNIT

 key until 

no

no

is displayed, then press the 

ON/TARE

key

to continue the setup.

LOAD WEIGHT

With the display will showing

LOd

LOd

, place a (2) lb certified test weight on the scale tray.  Allow 

approximately five seconds for the scale to stabilize.  Press the 

ON/TARE

key to start the 

calibration.  Dashes will be displayed from the left to the right of the display momentarily, then 
the display will change to 

uLd

uLd

.

OC

OC

CAL

CAL

Err

Err

ErA

ErA

Summary of Contents for Detecto PS-5

Page 1: ... DIGITAL PORTION SCALE PS 5 OWNER S Manual CARDINAL SCALE MFG CO 8528 M013 O1 Rev G PO BOX 151 WEBB CITY MO 64870 Printed in USA 09 00 PH 417 673 4631 FAX 417 673 5001 Web Site http www detectoscale com ...

Page 2: ...2 ...

Page 3: ...vidual accuracy skill and caution For this reason the Seller is not able to guarantee the result of any procedure contained herein Nor can they assume responsibility for any damage to property or injury to persons occasioned from the procedures Persons engaging the procedures do so entirely at their own risk TABLE OF CONTENTS Specifications 1 Installation 2 Battery Replacement 2 Operation 2 Automa...

Page 4: ...d batteries and replace with fresh ones Make certain the and ends of the batteries match the positions shown on the scale circuit boards as shown Replace the battery compartment cover and retaining screw OPERATION 1 TURN THE SCALE ON Press the ON TARE key 2 SELECT THE WEIGHING UNIT Press the WT UNIT key to alternate between ounces and gram weight displays A red indicator light will illuminate to s...

Page 5: ...h the drawing on the circuit board If using the optional AC power adapter insert the plug into the jack on the back of the scale BEGIN SETUP and CALIBRATION 1 Making certain the scale tray is empty press and hold the setup and calibration switch located on the lower right side of the scale Refer to the figure on the next page 2 Press the ON TARE key to turn the scale on Release the setup and calib...

Page 6: ... setting Press the WT UNIT key until the desired number of minutes of inactivity before the scale will automatically shutoff is displayed Values of 1 2 5 10 or 15 minutes or OFF are available Selecting OFF will disable the automatic shutoff feature Press the ON TARE key to save the setting POWER UP ZERO The display will show PuO PuO Press the WT UNIT key to display the current setting Press the WT...

Page 7: ...2 ...

Page 8: ...U N A M E D S I G U N E D F A C T R E D I N U S A U N A M E D S I G U N E D F A C T R E D I N U S A ...

Reviews: