INSTALLATION, CONT.
8.
Referring to the table below (or on the circuit board) for terminal connections, connect each
wire to terminal block P2 or P3. Refer to Figure No. 12 for terminal block location.
Load Cell Wiring Table
– P2 or P3
Board Label
Function
Board Label
Function
+EXC
+ EXCITATION
-SIG
- SIGNAL
+SEN
+ SENSE
-SEN
- SENSE
+SIG
+ SIGNAL
-EXC
- EXCITATION
SHLD
SHIELD (Not used, the shield wire is connected to metal gland).
9.
To terminate a wire, use a small flat blade screwdriver and press down on the release bar
for the terminal. Insert the wire into the opening and remove the screwdriver, allowing the
release bar to return to its original position, locking the wire in place. See Figure No. 9.
10.
Repeat the procedure until all wires are in place.
11.
After all terminations have been made, remove the excess cable from the enclosure.
3/4 inch
Figure No. 8
5 to 6 inches
1/4 inch
Insert
Wire
Press
Down
Figure No. 9