Configured Options (Continued)
Alaris® PK Editor Software - Pump Configuration
The following options are configurable via the Alaris® PK Editor Software (PC based), see the Alaris® PK Editor Software Directions for Use
(1000CH00016) for details on how to alter the profile configurations.
AC Fail Warning
The AC Power Failure Alarm can be set to sound or be silent if the AC power is disconnected.
Audio Volume
The audio alarm volume of the pump (high, medium or low).
Auto Night Mode
Main Display (Backlight) dims between hours 21:00 and 06:00.
Battery Icon
Indicator displaying the remaining estimated battery capacity.
Callback Time
Adjusts the length of time before the pump sounds the Call Back alarm.
Event Log
The Event Log can be set to be displayed on the main display. Events are still recorded in the
Event Log if disabled.
Drug Override Mode
Always
- Any changes made to the dose rate or target concentration that are outside the
editor Soft Alerts will require confirmation before starting infusion.
Smart
- Confirmation of setting will be required on the first dose rate or target concentration
set outside the editor Soft Alerts. Any subsequent changes will not require confirmation
until after the dose rate or target concentration has been confirmed inside the editor Soft
Alerts. Additionally, any changes in dose rate or target concentration from above a Soft
Alert Max to below a Soft Alert Min or from below a Soft Alert Min to above a Soft Alert Max
will also need to be confirmed.
Pressure Default
The default occlusion pressure alarm level.
Pressure Display
Sets whether the Pressure Information is available on the main display.
Purge Rate
The rate used during purge operation.
Purge Volume Max
The maximum permissible purge volume.
Purge Syringe Prompt
Feature which prompts the user to purge the extension set prior to the start of the infusion.
Bolus
1
Bolus feature can be set to HANDS ON or HANDS FREE.
Bolus Rate Default
1
The default bolus rate.
Bolus Volume Default
1
The default bolus volume.
KVO
Allows the enabling or disabling of Keep Vein Open (KVO) at End of Infusion (EOI).
KVO Rate
Sets the KVO rate at which the pump will operate when EOI is reached.
Near End of Infusion Time
Sets the Near End of Infusion warning time as time left to End of Infusion.
End of Infusion %
Sets the End of Infusion point as a percentage of syringe volume.
Weight Default
2
The patient default weight in kg.
Weight Minimum
2
The minimum patient weight in kg. This is a Soft Alert and can be overridden.
Weight Maximum
2
The maximum patient weight in kg. This is a Soft Alert and can be overridden.
Age Default
2
The default patient age in years.
Age Minimum
2
The minimum age in years. This is a Soft Alert and can be overridden.
Age Maximum
2
The maximum age in years. This is a Soft Alert and can be overridden.
A
The approved Data Set contains configurable option values per profile.
The originator and approvers of the Data Set should be aware that, unless a rationale for safety is provided, it is
not recommended to set the callback time to a value greater than the default setting of 2 minutes since doing so
would not be in compliance with IEC/EN60601-2-24:1998 standard.
1
The bolus configurations are used only when the Alaris® PK Syringe Pump is being used in ml/h mode. If a drug is selected then the drugs
own configuration settings are used.
2
Although a default and Soft Limits can be set for age and weight, the actual selectable range may be limited by the drug and model
chosen.