Cardiac Designs ECG Check Universal Manual Download Page 17

 
ECGC-70-002, Rev. 8, February 05, 2019 

 

17 

Troubleshooting 

 

Problem 

Possible Cause(s) 

Solution 

Device does not send ECG 
to application 

The battery is very low or 
depleted 

Re-charge battery 

 

The Bluetooth radio is not 
enabled on the cellular 
device 

Turn Bluetooth ON in the 
cellular device 

 

The ECG Check is out of 
range with the cellular 
device 

Move closer to the cellular 
device 

 the ECG Check 

device and the cellular 
device must be within 10 
meters for ECG data 
transfer. 

 

The user has not touched 
both electrodes. 

Ensure both electrode pads 
are in contact with the 
user

s skin 

ECG Data is not being 
analyzed 

There is no connection to 
the internet 

Turn on your cellular data 
or connect to a WiFi access 
point 

ECG analysis comes back 
with a bad recording 

Too much noise or 
movement in the 
recording 

Do not move & remain still 
while the device is 
recording 

 

Not enough ECG data was 
collected 

Keep the device in contact 
with the skin for 30 
seconds 

Cellular device is running 
slowly 

Too many applications 
open 

Manually close the 
applications or restart your 
cellular device. 

ECG Check application 
does not launch 

Too many applications 
open 

Manually close the 
applications not in use or 
restart your cellular device. 

 

 

 

Summary of Contents for ECG Check Universal

Page 1: ...or your safety please take a moment to read it carefully and in its entirety before you begin testing For the most up to date list of compatible devices please refer to the supported list at http www...

Page 2: ...the ECG Check 10 Setup the ECG Check Device 10 Download the ECG Check Application and Create your Account 10 General Information 11 Before Recording 11 Recording 11 Medical Records 14 Maintenance Clea...

Page 3: ...ng Radiation Authorized representative in the European Community Indicates the Authorized representative in the European Community European Conformity Serial Number 2498 Indicates the manufacturer s s...

Page 4: ...vided in this manual are subject to change without notice Cardiac Designs Inc SHALL NOT BE LIABLE FOR TECHNICAL OR EDITORIAL ERRORS OR OMISSIONS CONTAINED HEREIN NOR FOR INCIDENTAL OR CONSEQUENTIAL DA...

Page 5: ...nal defibrillation procedures WARNING The ECG Check device must not be exposed to strong electromagnetic fields WARNING Do not allow conductive parts of electrodes to come into contact with other cond...

Page 6: ...ould consult their physician before using this device Coronary heart disease Valvular heart disease Heart transplant Heart failure This device is available in the USA in both over the counter mode OTC...

Page 7: ...ing at home These 1 lead cardiac monitors allow remote patients to display and transmit their EGG data to medical professionals via a communication device to a remote server Specifically the EGG Check...

Page 8: ...ECGC 70 002 Rev 8 February 05 2019 8 Your ECG Check comes with the following ECG Check Device ECG Check USB Charging Cord...

Page 9: ...til the ECG Check application has been started and the blue Make a New Recording icon has been pressed 2 Standby Power Off The device will turn off automatically after the device has detected no skin...

Page 10: ...computer and store in an easy to find location so you will have it easily accessible for recharging when necessary Download the ECG Check Application and Create your Account 1 Download the ECG Check...

Page 11: ...ord and transmit your ECG to an application via Bluetooth for display and analysis as shown below Before Recording For your safety read thisinstruction manual in its entirety before attempting a recor...

Page 12: ...o be prompted on how to start recording 5 The following screen will appear once you have established a connection and the ECG Check device begins recording ECG Check Application Recording in Progress...

Page 13: ...w If you are in Over the Counter Mode you will only see a progress bar as the recording progresses 8 The system will automatically send the ECG for analysis by the backend arrhythmia analysis algorith...

Page 14: ...s 2 Tap on the Tap for archive recordings to view older recordings than those shown on the screen Maintenance Cleaning Rechargeable Battery Warning Do not dispose of the internal rechargeable battery...

Page 15: ...onthly at minimum to ensure continued performance of your ECG Checkdevice at maximum capacity and to reduce the possibility of a failure Refer to the cellular device instructions for proper maintenanc...

Page 16: ...id pouring water or other liquids on the device while cleaning Never use abrasives such as wire wool or metal polish Disinfection Using a new Sani Cloth HB cloth after cleaning wipe all exterior surfa...

Page 17: ...rodes Ensure both electrode pads are in contact with the user s skin ECG Data is not being analyzed There is no connection to the internet Turn on your cellular data or connect to a WiFi access point...

Page 18: ...Email info cardiacdesigns com Website www ecgcheck com Wireless Technology The ECG Check device system uses wireless technology to transmit data from the ECG Check Device to the ECG Check application...

Page 19: ...beats per minute measured as bpm Bluetooth BT Wireless communication protocol ECG Check ECG Check wireless ECG monitor PC Personal computer ECG Check Application Cardiac Designs Inc proprietary softw...

Page 20: ...ic and Medical Average QRS Sensitivity 99 87 Average QRS Predictability 99 77 Average VEB Sensitivity 93 91 Average VEB Predictability 94 12 Battery Type 3 7V x1 Lithium ion rechargeable battery Batte...

Page 21: ...that it is used in such an environment Emission test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The ECG Check uses RF energy only for its internal function Therefore...

Page 22: ...hospital environment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage dips short...

Page 23: ...ut power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determine...

Page 24: ...rs and the ECG Check as recommended below according to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency o...

Reviews: