1700-1800° Chamber 
 

 

Electrical Connections 

 

Connection by a qualified electrician is recommended.

 

 

The smaller models are designed only for single phase electrical supplies with or without neutral. 
The larger models are for 3-phase use, but can also be supplied for single phase. The furnace must 
be connected only to the type and voltage of supply for which it was ordered. 

 

The voltage or range of voltages on which the furnace may be operated is given on the furnace 
rating label. Check that the supply voltage is compatible with the voltage on the label, and that the 
current capacity is sufficient for the amperage on the label, before connection to the supply. A 
table of the most common ratings is given in section 8.1 of this manual. 

 

In some models supply cables are fitted, as are internal supply-rated fuses. The cable may be 
directly wired to an isolator or fitted with a line plug. 

 

The supply point must be within reach of the furnace operator and must incorporate either an 
isolating switch which operates on both conductors, or a quickly removable plug. 

 

The supply MUST incorporate an earth (ground). 

CONNECTION DETAILS 

supply type

           

 

Supply 

Terminal label 

Cable colour 

Live-Neutral 

Reversible or Live-Live 

1-phase

 

    L 

Brown 

To live 

to either power conductor 

 

    N 

Blue 

To neutral 

to the other power conductor 

 

    PE 

Green/Yellow 

To earth (ground) 

to earth (ground) 

supply 

Terminal label 

Cable colour 

 

3-phase

 

    L1 

Black 

to phase 1 

 

    L2 

Black 

to phase 2 

 

    L3 

Black 

to phase 3  

 

    N 

Light Blue 

to neutral  

except delta

 

 

    PE 

Green/Yellow 

to earth (ground) 

 
 

DO NOT connect a furnace ordered for three phase use to a single phase supply or to the wrong 
type of three phase supply. 

3.0 

MF22 

Summary of Contents for HTF 18/15

Page 1: ...nnected to the electricity supply CONTENTS Section page 1 0 Symbols Warnings 2 2 0 Installation 3 3 0 Operation 5 4 0 Maintenance 8 5 0 Repairs Replacements 8 6 0 Fault Analysis 12 7 0 Circuit Diagram...

Page 2: ...glows Heat Light the adjacent light glows or flashes to indicate that power is being supplied to the elements 1 2 Warning Symbols DANGER of electrical shock read any warning printed by this symbol DA...

Page 3: ...there is free space around the furnace Do not obstruct any of the vents in the case there are cooling fans in the case which must not be obstructed Ensure that the furnace is placed in such a way that...

Page 4: ...nual In some models supply cables are fitted as are internal supply rated fuses The cable may be directly wired to an isolator or fitted with a line plug The supply point must be within reach of the f...

Page 5: ...d Find and correct the cause before resetting the overtemperature controller according to the instructions supplied To switch the furnace off set the Instrument switch to off The case cooling fans rem...

Page 6: ...hearth can be protected from abrasion if required by the provision of a secondary plate This consumable item can either be in light weight ceramic fibre or in dense hard wearing alumina refractory 3...

Page 7: ...by a calibration test or by comparing the output with a new reference thermocouple which has been subjected to high temperatures for a minimum length of time Failure to check the thermocouple may res...

Page 8: ...ont of this booklet when supplied with a furnace To contact Thermal Engineering Services or Carbolite see the back page of this manual 4 4 Recommended Spares Kits Carbolite can supply individual spare...

Page 9: ...rately Before commencing any major repairs we recommend reference to the European Ceramic Fibre Industry Association Bulletin No 11 and the UK Health and Safety Executive Guidance Note EH46 We can pro...

Page 10: ...ponents and the support system appropriate to your furnace There are also variations in the positions of the braids which connect the elements the rule is that for single phase all elements are in ser...

Page 11: ...ler clips do no require clip tool Element Data distance from top of element clamp to top of element RHF 17 3 RHF 17 10 32mm RHF 17 25 HTF 18 3 47mm HTC 18 8 45mm HTF 18 15 40mm HTF 18 27 40mm Element...

Page 12: ...controller may be faulty or not receiving a supply due to a faulty switch or a wiring fault The SUPPLY light is OFF Check the supply fuses and any fuses in the furnace control compartment B Furnace O...

Page 13: ...ansformer may be fitted 2 Door actuator circuits may differ from the example shown 7 2 Three Phase Models Each set pair of elements has its own transformer controlled by a thyristor complete with thyr...

Page 14: ...1700 1800 Chamber 14 MF22...

Page 15: ...troller OP Hi should be 100 for all models The thyristor stacks are set to the secondary element maximum currents given in the table TABLE OF FUSE AND CURRENT LIMIT VALUES Model Phases volts supply fu...

Page 16: ...g the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specific...

Reviews: