23/04/2019
Rev 3
21
II.2.
Operating conditions
The gas analyzer is equipped with an automatic monitoring system of parameters that have an
influence on measurements. If at least one of these parameters is out of bounds, and might have
a metrological impact on results, then the gas analyzer automatically prevents any measurement
from being performed until operating conditions are stabilized.
Failure to observe operating conditions may lead to the equipment degradation or
to a temporary withdrawal of it:
Atmospheric pressure 1000 mbar +10% -25%
Power supply 230 VAC +10% -15% 50 Hz+-2%
Ambient temperature: from -10 to 55 °C
Gas temperature: 200 °C tolerated by the probe
Storage temperature: -32 to +55 °C
Relative humidity: <98% non-condensing
Ambient air clean and room well ventilated