HOOKUP CONNECTOR BLOCK LABEL
The Hookup Connector Block Label contains useful information for quick reference.
Handle Locking Instructions
- insert block onto basin channel manifold and move handle
to the left to lock. Move to the right to unlock.
Lot Number
- identifies date of manufacture.
Hookup Part Number
- identifies hookup.
Hookup Barcode
- barcode of part number.
Port Name
- identifier of port location in ADVANTAGE system. Error logs identify issues using
these names. Also identifier of Connector Assembly replacements for older model hookups.
Port ID Number
- identifier for Connector Assembly replacements. Not found on older
model hookups.
Alternative Channel Connector Names
- connector names vary depending on specific
endoscope channel configuration.
WARNINGS & CAUTIONS
This hookup is intended for connection only with the specific endoscope models
identified in the appropriate MEDIVATORS online hookup guide. Use of this hookup
with endoscopes other than those specified by the hookup application guide may
result in inadequate HLD.
Users operating MEDIVATORS AERs and hookups must be trained and competent in
the understanding of endoscope channel systems.
Prior to reprocessing, users must verify that the hookup used contains connections
for all appropriate channels that require a separate reprocessing connection
according to the hookup instructions
All connectors and adapters must remain firmly attached and unrestricted for the
entire disinfection cycle to ensure adequate HLD. Users must inspect the endoscope
and hookup combination to verify proper connection and flow.
Modifications or repairs to hookups which do not correspond to the manufacturer’s
specifications may result in inadequate HLD and/or damage to endoscopes.
Failure to properly clean and prepare endoscope for immersion prior to hookup
connection may result in inadequate HLD and/or damage to the endoscope.
Endoscopes must be inspected for damage and verified to be in proper working
order prior to hookup connection. Connection of hookups to damaged endoscopes
or endoscopes with obstructed/restricted channels could result in inadequate HLD.
Manual leak testing must be carried out in accordance with manufacturer’s
recommendations prior to reprocessing.
Install all necessary waterproof caps, plugs and cleaning adapters prior to
immersion.
Do NOT autoclave MEDIVATORS hookups.
MEDIVATORS makes no claim on high-level disinfection efficacy when these
reprocessing instructions are not followed, or when this hookup kit is applied to
endoscopes other than those specified in the online interactive hookup guide
.
Avoid biological contamination. Always wear personal protective equipment when
handling endoscopes.
© 2016 Medivators Inc. 50098-562/B
CASSETTE-STYLE HOOKUPS (not available in US)
An ADVANTAGE Hookup Cassette combines an endoscope tray with the hookup connector
block. Specific endoscope connections for this hookup are identical to non-cassette
versions. Cassette-style hookup part numbers are designated with a CAS suffix.
ADVANTAGE Hookup Cassettes are available outside the United States only. Connector
Assemblies are listed in the Connector Assembly Identification Chart.
LO
T:
XXXXXX
ID
:
XXXX
P/N: X-X-XXXHAN
Hookup Part Number
Hookup
Barcode
Port ID
Number
Port Name
Alternative
Channel
Connector
Names
Handle Locking
Instructions
Lot Number
ADVANTAGE
™
and MEDIVATORS
®
are trademarks and registered
trademarks of Medivators Inc.