Locate the endoscope port and follow the instructions for attaching the
corresponding Endoscope Connector. All connectors must be attached.
Air/Water Connector
Align the connector with
the air/water bottle port and
push to fully engage.
2-8-650HAN/2-8-650CAS
HOOKUP CONNECTION GUIDE
Using MEDIVATORS
®
ADVANTAGE™ Automated Endoscope Reprocessors
Leak Test Connector
Align the slot on connector with the
pin on the scope. Push on and turn
clockwise to engage. To remove
leak test connector, turn connector
counter-clockwise until slot is
aligned with pin and pull off. Do not
pull until slot is aligned with pin.
READ ALL INSTRUCTIONS PRIOR TO CONNECTION
Port ID
Port Name
Connector Assembly Description
Order Number
1
LEAK TEST
Leak Test Connector Assembly
H28650-1
2
JET/AUX
Air/Water Connector Assembly
H28650-26
3
BIOPSY
Biopsy Inlet Connector Assembly
H28650-3
4
AIR
Air Probe Connector Assembly
H28650-4
5
ELEVATOR
Forward Water Jet Connector Assembly
H28650-5
6
WATER
Air/Water Connector Assembly
H28650-26
7
SUCTION
Suction Connector Assembly
H28650-7
A
NA
Air/Water, Suction Channel Plug
N/A
If a Connector Assembly needs replacement, locate the port number/name and use
the corresponding order number found in the Connector Assembly Identification Chart.
Connector Assemblies are specific to individual hookups. Check each Hookup Connection
Guide for relevant information.
Each Advantage Hookup is comprised of multiple ports and connector assemblies. Each
port is labeled with a Port ID number (circled). This same number also identifies the
corresponding Connector Assembly. Each Connector Assembly is fitted with a specific
endoscope connector. (see Fig. 2).
EXAMPLE CONNECTOR ASSEMBLY
(Figure 2)
CONNECTOR ASSEMBLY IDENTIFICATION CHART
Ferrule
Endoscope
Connector
example
Tubing
Connector
Coupling
Label
ADVANTAGE HOOKUP AND CHANNEL SEPARATOR
(Figure 1)
Biopsy Inlet Connector
Align the biopsy connector
disc over the biopsy inlet
and push down to compress
spring and slide the frame
forward to fully engage.
Air Probe Connector
Align with air probe, push
and slide until fully engaged.
Suction Connector
Align silicone tubing with suction
barb and push tubing onto barb
until fully engaged.
Air/Water, Suction
Channel Plug
Align and attach Peer
Medical channel plug
(supplied with endoscope)
to air/water and suction
ports. Insert and close
clamp to secure. Ensure
that luer caps are fitted
and secure.
A
Forward Water Jet Connector
Align stainless steel fitting with forward
water jet port, push and turn clockwise
to connect.
LO
T
:
XXXXXX
ID:
xxxX
P/N: X-X-XXXHAN
Ensure proper alignment of all connectors with the
endoscope mating parts prior to attachment.
If you have any questions regarding this hookup, in the U.S., please contact MEDIVATORS Technical Support at
1-800-444-4729. Outside the U.S., please contact your MEDIVATORS Distributor.
DESIGNATED HOOKUP FOR THE FOLLOWING
ENDOCHOICE
®
ENDOSCOPES*:
C38s
C38s-133
C38s-150
CS-1001
F1C-133
F1C-150
FUSE-1C
FUSE-1G
FUSE-F1C
FUSE-G32
* This list of corresponding endoscope models is subject to change without notice. For a full list, please go to:
www.medivators.com/hookuplookup
© 2016 Medivators Inc. 50098-562/B
Representative endoscope for illustrative purposes only. For exact
.
port locations, refer to endoscope manufacturer’s instruction guide.
Forward Water Jet Cleaning
Adapter Gas 2426
This is an optional piece
Air/Water, Suction
Channel Plug Gas 2393
ADDITIONAL REQUIRED ENDOSCOPE CONNECTIONS