background image

Operating procedures

15

To set the EIt value

Elt value can be set in the processing parameter edit screen of the CXDI Control Software.
Enter a value in the EIt text box referring to the current EI value.

• EI (Exposure Index)

EI is an approximative indicator of the dose that reaches the detector, as calculated per IEC 
62494-1 standards from the captured images.

• EIt (Target Exposure Index)

EIt is the target EI level.

• DI (Deviation Index)

DI is a logarithmic index indicating the difference between the EI and EIt.

3.8

Image processing adjustment

000_BT8-1855-EN01_CXDI-Pro.book  Page 15  Tuesday, November 9, 2021  5:14 PM

Summary of Contents for CXDI-Pro

Page 1: ...DIGITAL RADIOGRAPHY Instructions for Use CXDI Pro 000_BT8 1855 EN01_CXDI Pro book Page 1 Tuesday November 9 2021 5 14 PM ...

Page 2: ...1 IEC 60950 1 or IEC 62368 1 For further details consult your sales representatives or local Canon dealers 5 Connection of a system that uses the product to a network could result risks to patients operators or third parties A dedicated individual who is assigned responsibility for maintenance should assess these risks in advance The responsible person should also assess the risks when changes to ...

Page 3: ...r where you purchased the product The information above including information on batteries is on our website in the official languages of each EU country Please access https global canon en ifu medcom envfile weee battery eu pdf Only for the United Kingdom This symbol indicates that this product is not to be disposed of with your household waste according to the UK Waste Electrical and Electronic ...

Page 4: ...or fire or injury Close to facilities where water is used Where it will be exposed to direct sunlight Close to the air outlet of an air conditioner or ventilation equipment Close to a heat source such as a heater Where the power supply is unstable On the floor In a dusty environment In a saline or sulfurous environment Where temperature or humidity is high Where there is freezing or condensation I...

Page 5: ...osing patients with cardiac pacemakers to X rays always make sure that the patients have no problems during exposure If a problem occurs keep the equipment wiring cable or PC Connection Cable away from the patients and consult your doctor Always connect the three core power cord plug to a grounded AC power outlet To make it easy to disconnect the plug at any time avoid putting any obstacles near t...

Page 6: ...following precautions before using the equipment The ingress of water may damage the equipment and cause a fire or electric shock Do not submerge the detector in water If the detector gets wet use a dry soft cloth to wipe it completely dry Securely close and lock the battery cover The dust proofing and waterproofing performance may be compromised if the cover is not locked Do not open or close the...

Page 7: ...kin or clothes immediately wash it off with tap water etc The contact with the battery liquid may cause skin irritation While preparing for examinations be sure to confirm that the entered information patient name ID number date of birth and sex matches that of the patient If the information is incorrect the resulting patient mix up and a misdiagnosis may cause harm to the patient Be sure to use t...

Page 8: ...atient contact surfaces of the detector using a disinfectant such as disinfecting ethanol to prevent the risk of infection A blood infection or other causes may result in the onset of an infectious disease For details on how to disinfect consult a specialist Do not spray the detector directly with disinfectants or detergents Always keep this product and other equipment clean and remove all dust an...

Page 9: ...his product may be affected by radio interference from other devices that generate radio waves microwave ovens Bluetooth devices digital cordless phones etc Keep the product as far away from those devices as possible to avoid radio interference during use This product is suitable for use in hospital professional healthcare facility environments with the exception of environments near active HF SUR...

Page 10: ...re is performed on a patient Images may not be acquired or an artifact may appear depending on the exposure conditions X ray dose irradiation time irradiation field or target body positioning even if an exposure is performed Images and raw data stored on the recording medium in the equipment may become unreadable due to operation errors equipment failure or other unexpected events so be sure to al...

Page 11: ...place radiographic film screen systems in all general purpose diagnostic procedures This device is not intended for mammography applications The image is obtained as the result of imaging X rays passed through the human body with an X ray flat panel detector and importing a digital signal output from the detector into the image processor 000_BT8 1855 EN01_CXDI Pro book Page 11 Tuesday November 9 2...

Page 12: ... turn it on 3 Connect the detector to the image capture computer 1 Enter the required items in the control software such as the Patient ID and target body part 2 After confirming that the ready indication is displayed on the monitor of the image capture computer press the exposure switch of the X ray generator to irradiate the patient with X rays 3 1 Preparations System startup Preparing to use th...

Page 13: ...essing takes place using the image processing parameters set beforehand by the operator 2 Metadata such as Patient ID information is assigned to the image data following image processing and the image information is transferred to the specified device 1 Shut down the image capture computer 2 Press the POWER switch of the detector to turn it off 3 3 Checking exposed images 3 4 Data processing 3 5 E...

Page 14: ...d Flashing slowly in 4 second cycles max 8 seconds 2 Linkage ended Illuminated Illuminated 2 sec Switching to exposure ready status Illuminated Flashing Exposure ready status Illuminated Illuminated Detector selection status Sleep Illuminated Not illuminated Error Flashing Flashing 3 7 List of lights flashes notification beeps of the ready indicator Link registration completed During exposure read...

Page 15: ... EI Exposure Index EI is an approximative indicator of the dose that reaches the detector as calculated per IEC 62494 1 standards from the captured images EIt Target Exposure Index EIt is the target EI level DI Deviation Index DI is a logarithmic index indicating the difference between the EI and EIt 3 8 Image processing adjustment 000_BT8 1855 EN01_CXDI Pro book Page 15 Tuesday November 9 2021 5 ...

Page 16: ...rsists turn off the detector consult your sales representative or local Canon dealer and inform them of the details of the problem error codes or error messages Symptom Cause Error messages in the CXDI Control Software Remedy Power LED lamps flash 1 second cycles and exposure is not possible No battery power Replace the battery pack with a fully charged one and retry an exposure 000_BT8 1855 EN01_...

Page 17: ... missing screws 2 Ensure that there are no break or no deformation on the exterior of the detector 3 Ensure that there is no dust or foreign matter on the battery bay connector 4 Ensure that there are no breaks or short circuits in the battery bay connector 5 Ensure that there is no dust dirt or oil on the terminals of the cable connector 6 Ensure that the battery cover is not damaged or bent 7 En...

Page 18: ...etc 2 Check the EI value captured 1 Perform calibration in the following circumstances When exposure conditions have changed significantly When images appear strange in some way When the installation environment has changed significantly Monthly inspection Yearly inspection Irregular inspection 000_BT8 1855 EN01_CXDI Pro book Page 18 Tuesday November 9 2021 5 14 PM ...

Page 19: ... 10 to 95 RH without condensation 613 to 1060 hPa Applicable grid Other devices 40 52 lp cm recommended Reducing scattered radiation Even when a grid is not attached to the detector the reduction in contrast caused by scattered radiation can be improved by image processing Adjust the power of scatter correction by selecting the Effect check box to enable this control on the Scatter Correction cont...

Page 20: ...interfaces Card reader Connection interface USB Readable cards JIS X6301 1998 compliant Readable encoding JIS II Power supply USB bus power Barcode reader Connection interface USB Compatible encodings Code39 Code93 Code128 JAN EAN 8 JAN EAN 13 Industrial 2 of 5 Interleaved 2 of 5 Matrix 2 of 5 MSI NW 7 UPC A UPC E RSS EAN 128 Plessey PDF417 Power supply USB bus power USB cable Connector micro B co...

Page 21: ...QE value at 3 5 μGy in 0 5 lp mm RQA5 is 0 58 with measurement error of less than 10 Access point Use a product that satisfies the following requirements Supports IEEE 802 11a b g n ac Supports WPA2 or WPA3 Supports 10 100 1000BASE T Ethernet network interface 6 2 Characteristics 000_BT8 1855 EN01_CXDI Pro book Page 21 Tuesday November 9 2021 5 14 PM ...

Page 22: ...ps global canon en ifu medcom index html For details please contact your sales representative or local Canon dealer 7 1 Medical equipment classification Type of protection against electrical shock Internally powered equipment Class I Equipment with Power Box Degree of protection against electrical shock Type B Applied Parts Detector 7 2 CE marking Regulation EU 2017 745 Directive 2011 65 EU Direct...

Page 23: ...his mark indicates general precautions Consult instructions for use Non ionized radiation Manufacturer Date of manufacture Serial number This mark shows a certification mark in European Economic Area This mark indicates that this equipment must be collected separately under the Directive on Waste Electrical and Electronic Equipment WEEE in the European Union This mark indicates medical device whic...

Page 24: ...loop fastener fastener hooks fastener loops One pair is already applied to the product 3 pairs PC Connection Cable CP 01 Wiring cable WC 01 Power Box PB 01 Power Box 1 AC adapter 1 Power cord 1 X ray Interface Unit XB 1A X ray interface box 1 AC adapter 1 Power cord 1 X ray interface cable 1 Functional earth conductor 1 Battery Charger BC 1A Battery charger 1 Power cord 1 CXDI Pro system 000_BT8 1...

Page 25: ...es 3 1 Improve the reduction in contrast caused by scattered radiation 2 Enhance the display of catheters bony parts such foreign matter as gauze etc in a captured image 3 Images can be rotated one degree at a time Optional function of software 000_BT8 1855 EN01_CXDI Pro book Page 25 Tuesday November 9 2021 5 14 PM ...

Page 26: ...this product will be stocked for 8 years after discontinuance of production to allow for repair The following consumable can deteriorate because of its characteristics and structure For purchase of consumables contact your sales representative or local Canon dealer Battery Pack LB 4A Cycle life Approx 300 cycles Battery cover Replacement period Approx every 2 years For the technical description se...

Page 27: ...000_BT8 1855 EN01_CXDI Pro book Page 27 Tuesday November 9 2021 5 14 PM ...

Page 28: ...maikami cho Nakahara ku Kawasaki Kanagawa 211 8501 Japan Telephone 81 3 3758 2111 Revision Date 2021 11 BT8 1855 EN01 1121P0 001 CANON INC 2021 BT8 1855 EN01 000_BT8 1855 EN01_CXDI Pro book Page 28 Tuesday November 9 2021 5 14 PM ...

Reviews: