4
Operating procedures
33
2
Register the detector and make connection to the control
system.
Press and release the POWER switch
again.
READY LED flashes
(approx. 3 seconds).
While READY LED flashes (approx.
3 seconds), locate the IR data port of
the detector near the ready indicator
or the IR data communication unit
(within 30 cm).
When the registration is completed, the
two LED lamps on the ready indicator
flash three times for 0.5 seconds at 0.5
second intervals, and a three-tone beep
is emitted.
Proceed to the next
Connection
.
Communication link is automatically
established between the internal
wireless module of the detector and
the wireless access point/control
system.
3
Conduct examination.
Registration
Connection
To reduce exposure to RF energy during examination, keep hands and other
body parts out of close contact with the antenna.
Important
000_DIA-2019-001-E04_AX-B3543W.book Page 33 Friday, November 27, 2020 12:25 PM