| CAV 4.2 Operation Manual
Run a Sample Test (
CAV 4.2 GUI
)
This test verifies system calibration and provides a technician with basic training in operating the
CANNON CAV 4.2. See Table 14 for tools and materials required.
Table 14 — Tools and Materials Required to Run a Sample Test
Description
Description
Assembled CAV 4.2
Sample test oil
Sample vials
Sample preparation materials
1.
Following your laboratory practices, prepare a 10 mL sample of test oil in the provided sample vials,
and then insert the sample vial into position 1 of the appropriate carousel.
2.
From the main screen, click the circle representing position 1 on the appropriate bath. The methods
screen opens. See Figure 16.
Figure 16 — Methods Screen
3.
Click
Verify Known KV
. The field entry screen open. See Figure 17.
Figure 17 — Verify Known KV Field Entry Screen
4.
Enter the sample ID, then click the green checkmark to submit it. Another field entry screen opens.