New GentleYAG

 Installation Procedure 

 Candela Corporation 

 

 

8503-01-0831,Revision B 

Candela Corporation Proprietary    

Page5 of 6 

 

  

A-  Handpiece Control Connector 

B-  Cryogen Output Connector 

C-  Fiber Proximal End (should be blue) 

D-  Glass Windows   

E-  Distance Gauge  

F-  O Ring 

     

 

          E                      D            F 

Figure 2. Distance Gauge  

5.  Distance Gauge Installation: 

5.1. Refer to figure 1 above for distance gauge location. 

Note: The distance gauge must be removed before performing the calibration 
procedure. 

5.2. Insert the distance gauge into the distal end of the handpiece delivery 

system. 

6.  Footswitch Installation: 

6.1. Remove the footswitch from its package. 

6.2. Attach the tubing end to the bulkhead connector on the rear panel of the 

laser. 

7.  COMPLETE THE DATA SHEET, 10-400-00145, PER THE CALIBRATION AND 

VERIFICATION DATA SHEET, 8503-01-0832. 

8.  Relocation: 

8.1. Care should always be taken when moving the 

GentleYAG.

  The system was 

designed to be moved, but special care should be taken when thresholds, 
elevator doors, ramps, and other uneven or sloping floor surfaces are 
encountered.  A severe physical shock could cause the alignment of the 
laser head or the optical fiber to be disturbed.  Furthermore, if the system is 
allowed to get out of control when being m oved, personal injury or physical 
damage could result. 

8.2. If it becomes necessary to relocate the 

GentleYAG

, call Candela or the 

distributor for details.  Failure to do so may result in damage to the system, 
and may void any warranty.  