New GentleYAG
Installation Procedure
Candela Corporation
8503-01-0831,Revision B
Candela Corporation Proprietary
Page4 of 6
4. Delivery System Installation:
4.1. Remove the Delivery System from its box.
4.2. With the system turned off, place the distal end of the handpiece assembly into
the calport.
4.3. Connect the handpiece control (round, electrical plug) connector to the
handpiece valve control receptacle on the front of the laser system .
4.4. Connect the cryogen output connector to the cryogen line receptacle on the front
of the laser system: Using both hands, push the knurled connector on the front
of the laser system away from you while gently pushing th e mating connector
into the knurled connector until it stops. Release the knurled connector.
4.5. Remove the protective cap on the fiber receptacle located on the front of the
system.
4.6. Remove the cap from the distal end of the fiber. Slide the distal end of the fiber
into the HP Slider until it clicks firmly into place.
4.7. Insert the proximal fiber connector into the proximal fiber receptacle, located on
the front of the system, until it clicks into place and stops.
Insert
Distance
Gauge
Here
Electrical
Connector
DCD
Connector
Fiber End
(Pugs into
Laser
Aperture)
Figure 1: Delivery System and Handpiece Assembly