6. In Output Setup, type the scan rate, meaningful table names, and Data Output Storage
Interval.
CAUTION:
The scan rate must be evenly divisible into fifteen minutes or one hour (depending on
the Measure every selected in step 3). The 0.5 m probe can take one minute, and the
1 m probe can take one and a half minutes to complete measurements. If the scan rate
is less than that length of time, skipped scans will be registered by the data logger.
7. Select the measurement and its associated output option.
SoilVUE™10 Complete Soil Profiler
4