background image

Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe 

8.  Calibration 

The multiplier is used to calibrate the CS512 probe.  To calculate the multiplier: 

1)

 

Program the datalogger using a multiplier of one. 

2)

 

Place the OxyGuard probe in the air, shaded from the sun.  Wait for readings 
to stabilize.  This may take 15 minutes or more. 

3)

 

Determine the air temperature and barometric pressure. 

4)

 

Using a calibration chart such as that provided in the probe’s manual, 
determine the oxygen concentration of the air. 

5)

 

Use the following equation to calculate the multiplier: 
 

M = P/R 

 

M = Multiplier 

 

P = Concentration in PPM of the air (

from the calibration chart in Appendix A

 

R = The signal output of the OxyGuard probe when using a multiplier of one 

6)

 

Change the multiplier in the datalogger program from one to the calculated 
number. 

A more common way to enter the multiplier is to insert a separate instruction in 
the program.  This will allow a new multiplier to be added to the program 
without rewriting, compiling, and downloading the program to the data logger. 

In the CR510, CR10X, and CR23X, use Instruction P36.  The multiplier is 
entered into an input location called DOmult using the numeric display in 
PC200W, PC208W, LoggerNet, PC400, PConnect, PConnectCE, or the 
CR10KD. 

EXAMPLE 6.  Sample Program using P36 

57:  Z=X*Y (P36) 
  1:  1 

  X Loc [ DOmV         ] 

  2:  2 

  Y Loc [ DOmult       ] 

  3:  3 

  Z Loc [ DOppm        ] 

 
In all dataloggers that use CRBasic, an expression is written.  The multiplier 
value is entered into the expression through the Public Table using the numeric 
display in PC200W, LoggerNet, PC400, PConnect, and PConnectCE. 

EXAMPLE 7.  Sample Program using Expression 

DOppm = DOMult * DOmV 

 

Summary of Contents for OXYGUARD TYPE III CS512

Page 1: ...Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe Revision 12 06 C o p y r i g h t 2 0 0 1 2 0 0 6 C a m p b e l l S c i e n t i f i c I n c...

Page 2: ...u of all other warranties expressed or implied including warranties of merchantability or fitness for a particular purpose CAMPBELL SCIENTIFIC INC is not liable for special indirect incidental or cons...

Page 3: ...formation 1 2 Specifications 2 3 Optional Probe Accessories 3 4 Optional Agitator 3 5 Application Information 4 6 Wiring 4 7 Programming 5 8 Calibration 6 9 Maintenance 7 10 Agitator Control 7 Appendi...

Page 4: ...P2 5 3 Sample Program using VoltSE 5 4 Sample Program using VoltDiff 5 5 Sample Program using VoltSE 5 6 Sample Program using P36 6 7 Sample Program using Expression 6 8 Sample Program using P86 and...

Page 5: ...the probe produces a millivolt output proportional to the oxygen present in the medium in which it is placed The probe consists of two parts an upper part with cathode anode and cable and a lower par...

Page 6: ...rin Membrane O ring Buna N Body Seal O ring Viton 0 05 mm 2 mil Teflon Dimensions and Weight 2 28 diameter x 2 3 height 15 9 oz Cable standard length 7 m 23 ft other lengths available upon request 2 w...

Page 7: ...es the water also has a high bio loading and the probes become fouled resulting in inaccurate DO measurements FIGURE 4 1 Preventing Bio fouling of the CS512 The PT4 Agitator overcomes these problems T...

Page 8: ...plication Information SCWin users This manual was written primarily for those whose needs are not met by SCWin Your procedure is much simpler just add the CS512 probe it s in the water folder save you...

Page 9: ...60 Hz Rejection Range code 23 used for CR23X 3 1 SE Channel 4 1 Loc DOmV 5 1 0 Multiplier See Calibration 6 0 0 Offset EXAMPLE 2 Sample Program using P2 1 Volt Diff P2 1 1 Reps 2 24 250 mV 60 Hz Reje...

Page 10: ...ing a multiplier of one 6 Change the multiplier in the datalogger program from one to the calculated number A more common way to enter the multiplier is to insert a separate instruction in the program...

Page 11: ...peat cycle timer requires no datalogger programming However some users choose to use a solid state relay and have the datalogger agitate the water on the probe face either periodically throughout the...

Page 12: ...et instruction EXAMPLE 9 Sample Program using Portset Portset 1 1 Delay 500 msec Portset 1 0 In the CR1000 CR800 CR850 and CR3000 dataloggers use the Portset instruction EXAMPLE 10 Sample Program usin...

Page 13: ...A 1 Appendix A Dissolved Oxygen in Fresh Water Table...

Page 14: ...Appendix A Dissolved Oxygen in Fresh Water Table A 2...

Page 15: ...This is a blank page...

Page 16: ...campbellsci com br suporte campbellsci com br Campbell Scientific Canada Corp CSC 11564 149th Street NW Edmonton Alberta T5M 1W7 CANADA www campbellsci ca dataloggers campbellsci ca Campbell Scientif...

Reviews: