SECTION 3.  TROUBLESHOOTING

3-3

The line monitor is connected to the RS232
serial data cable between the computer and
short haul modem.  The data line monitor has
seven lights indicating the status of the RS232
serial communication.  By observing these lights
in the order given below you may be able to
determine the problems with the computer link.

Pin 6 - DSR

Data Set Ready.  The short
haul modem holds this line high
all the time.

Pin 20 - DTR

Data Terminal Ready.  The
software holds this line high.

Pin 4 - RTS

Request to Send.  When the
computer tries to call the
weather station, it brings the
RTS line high until the call is
answered, or until the computer
stops trying to call.

Pin 5 - CTS

Clear To Send.

Pin 8 - CD

Carrier Detect.  Both CTS and
CD lines come high when the
calling modem receives a
signal indicating that the
weather station answered the
call.  If these lines do not come
high, then the communication
cable and wiring connections
should be double checked.

Pin 2 - TD

Transmit Data.  When the
computer sends commands to
the weather station, the TD light
will flash on.  This indicates that
the computer is transmitting out
data.

Pin 3 - RD

Receive Data.  The weather
station echoes back responses
to the commands that the
computer has sent.  The RD
light flash as these responses
are coming in.

3.2  SENSORS

Whenever sensor failure is suspected, the
sensors can be checked with a VOM
(Volt/Ohm meter) to measure for open
circuits (Section 3.2.1), or the CR10KD
(Keyboard Display) can be connected to the
CR10 to check sensor reading values
(Section 3.3.1 & Table 2.1).

3.2.1 OHM METER TESTS OF SENSORS

All sensors that are to be checked, EXCEPT for
the rain gage, should have a measurable
resistance.  INFINITE resistance indicates an
open circuit.  A sensor with an open circuit
should be sent in for repair.

The sensors which can be checked with an ohm
meter are:

• 

Anemometer (Wind Speed)

• 

Wind Direction Indicator

• 

Pyranometer (Solar Radiation)

• 

Rain Gage

• 

Temperature

NOTE: The measurable resistance of each
sensor is NOT intended to determine the
accuracy of the sensor.

For Ohm meter tests of the sensors refer to
sensor schematics in the appendix section.

Wind Speed

Check coil resistance between the black and
clear wires for an OPEN circuit.  The coil
resistance should measure between 1200 to
1300 Ohms.

Shorts to Ground - Check the base of the
sensor for moisture, or for any wires that may
have been pinched and exposed.

Wind Direction

The wind direction sensor varies its resistance
as the vane turns.  With the Ohm meter
connected to the black and red wires, the
resistance should change from 1K Ohms to 11K
Ohms with a five degree open or dead band.

If infinite resistance is measured, the sensor
needs to be repaired.

With the Ohm meter connected to the black and
white wires, the resistance should read 10K
Ohms.  If infinite resistance is measured, then
the potentiometer is bad and should be
replaced.

Solar Radiation

The Solar Radiation Sensor should measure
between 60 Ohms and 100 Ohms.  An open
circuit should be the main thing to look for.

Summary of Contents for NW8002

Page 1: ...NW8002 WEATHER STATION INSTALLATION INSTRUCTION MANUAL REVISION 3 97 COPYRIGHT c 1991 1997 CAMPBELL SCIENTIFIC INC...

Page 2: ...antability or fitness for a particular purpose CAMPBELL SCIENTIFIC INC is not liable for special indirect incidental or consequential damages Products may not be returned without prior authorization T...

Page 3: ...HOOTING 3 1 3 1 Communication Testing 3 1 3 2 Sensors 3 3 3 3 Datalogger 3 5 FIGURES 1 2 1 Effect of Structure on Wind Flow 1 2 1 3 1 NW8002 Enclosure with RAD Connection 1 3 1 3 2 NW8002AC SP Weather...

Page 4: ...Probe Assembly 2 3 3 1 1 RAD Testing Computer End 3 1 3 1 2 RAD Testing Datalogger End 3 2 3 2 1 4831 Temperature Relative Humidity Probe Assembly 3 5 TABLES 1 2 1 Approximate Range Miles 1 1 1 5 1 P...

Page 5: ...0 C Operating relative humidity range 0 to 100 non condensing SENSOR SPECIFICATIONS Temperature Range 35 C to 48 C Accuracy 4 C typically better than 2 C Relative Humidity Range 12 97 RH Accuracy 5 Wi...

Page 6: ...anted earth This allows the temperature and humidity readings to closely correspond with that of the turf the ET value is to represent 3 The radiation shield that covers the probe provides protection...

Page 7: ...ne modem for use with Mir 5000 systems only Figure 1 9 1 PS12 LA 12 V Rechargeable Battery Sensors Figure 1 3 2 4776 Li Cor Solar Radiation Sensor 6575 HMP35C Temperature and RH Probe 4830 Texas Elect...

Page 8: ...SECTION 1 WEATHER STATION DESCRIPTION AND INSTALLATION 1 3 FIGURE 1 3 1 NW8002 Enclosure with RAD Connection...

Page 9: ...SECTION 1 WEATHER STATION DESCRIPTION AND INSTALLATION 1 4 FIGURE 1 3 2 NW8002AC SP Weather Station Cross Arm Detail...

Page 10: ...SECTION 1 WEATHER STATION DESCRIPTION AND INSTALLATION 1 5 FIGURE 1 3 3 NW8002AC Weather Station...

Page 11: ...ft Channel up Step 3 Mount the wind set on the left side Mount the wind set so that the wind speed is on the side closest to you Step 4 Mount the solar radiation and rain gage sensors on the opposite...

Page 12: ...1 7 FIGURE 1 4 1 NW8002 Cement Base FIGURE 1 4 2 NW8002 Tower with Tilt Base...

Page 13: ...SECTION 1 WEATHER STATION DESCRIPTION AND INSTALLATION 1 8 FIGURE 1 4 3 NW8002AC SP Weather Station Top View FIGURE 1 4 4 Plumbing NW8002 Tower...

Page 14: ...mits current from the battery and provides circuitry to connect an external 12 volt battery The terminals on the PS12 are exposed by unscrewing the two set screws as shown in Figure 1 5 1 The two 12 v...

Page 15: ...gas may be generated at a rate sufficient to create a hazard Campbell Scientific makes the following recommendations 1 A CR10 equipped with standard lead acid batteries should NEVER be used in applic...

Page 16: ...SECTION 1 WEATHER STATION DESCRIPTION AND INSTALLATION 1 11 FIGURE 1 6 1 NW8002SP Weather Station...

Page 17: ...SECTION 1 WEATHER STATION DESCRIPTION AND INSTALLATION 1 12 FIGURE 1 6 2 NW8002 Weather Station Sensor Connection...

Page 18: ...modems Installation is as follows 1 Select a direct burial cable with two twisted pairs Two recommended types of cable are Two Cables Toro shielded armored cable P 7162 D A 1 or Anixter F 02P22BPN Rod...

Page 19: ...SECTION 1 WEATHER STATION DESCRIPTION AND INSTALLATION 1 14 FIGURE 1 8 1 RAD Modem Connection and Grounding...

Page 20: ...the computer site to communicate with the weather station The DC112 comes from the factory connected to the datalogger inside the enclosure Figure 1 9 1 The burial phone cable is routed through the b...

Page 21: ...SECTION 1 WEATHER STATION DESCRIPTION AND INSTALLATION 1 16 FIGURE 1 9 1 NW8002 Enclosure with DC112 Phone Modem Option Use with MIR 5000 Systems Only...

Page 22: ...2 1 Look for possible cable damage on any sensors and record all inspection findings in log section 2 4 2 2 MONTHLY 2 2 1 RAIN GAGE Check the rain gage funnel and tip mechanism Be sure screen inside f...

Page 23: ...pes of Temperature Relative Humidity Sensors which you may have Refer to the cleaning procedures for your particular system below 2 3 2 TEMPERATURE RELATIVE HUMIDITY SYSTEMS SHIPPED AFTER JANUARY 1 19...

Page 24: ...screen off and thoroughly clean the screen DO NOT TOUCH THE RELATIVE HUMIDITY CHIP Remember that while the sensor is rugged the function of it is to sense water vapor in the air therefore careful han...

Page 25: ..._ _____ ____________________________ _____ ____________________________ _____ ____________________________ _____ ____________________________ _____ ____________________________ _____ _________________...

Page 26: ...____________________ _____ ____________________________ _____ ____________________________ _____ ____________________________ 2 TO 3 MONTH Rain Gage Inspect and clean if necessary Date OK Comments ___...

Page 27: ...spect and spin each sensor Date OK Comments _____ ____________________________ _____ ____________________________ _____ ____________________________ _____ ____________________________ _____ __________...

Page 28: ...sor leads for damage or corrosion Also check leads at terminal strip of datalogger to be sure that leads are securely fastened b Use circuit diagrams in Appendix and Section 2 3 to ohm out defective s...

Page 29: ...ill has not been established go to Sections 3 1 thru 3 3 to test the datalogger If the datalogger test passes then the SC932 is suspect and will need to be repaired or replaced 3 1 2 TESTING THE SC95C...

Page 30: ...circuits Section 3 2 1 or the CR10KD Keyboard Display can be connected to the CR10 to check sensor reading values Section 3 3 1 Table 2 1 3 2 1 OHM METER TESTS OF SENSORS All sensors that are to be c...

Page 31: ...f you are suspecting improper readings and if facilities exist the temperature and relative humidity probe should be placed in a sealed container over deionized water for two hours After two hours it...

Page 32: ...OTE Contact your Toro distributer for RH chip replacement 3 3 DATALOGGER CR10 The CR10 is the heart of the weather station it measures the sensors outputs and converts the measurements to engineering...

Page 33: ...CR10 Temperature and 10 Battery Voltage If the datalogger is not operating properly the CR10 Temperature will not record a reasonable value The battery voltage should read no less than 11 76 volts If...

Page 34: ...A 1 APPENDIX A SENSOR SCHEMATICS...

Page 35: ...APPENDIX A SENSOR SCHEMATICS A 2...

Page 36: ...APPENDIX A SENSOR SCHEMATICS A 3...

Page 37: ...APPENDIX A SENSOR SCHEMATICS A 4...

Page 38: ...APPENDIX A SENSOR SCHEMATICS A 5...

Page 39: ...APPENDIX A SENSOR SCHEMATICS A 6...

Reviews: