a. Double click on the value in the field that needs to be changed:
b. Enter the correct value:
c. Press Enter to set the correct value.
4. When all values are corrected, set the ApplyandRestart variable to 1 and press Enter to reset
the system and apply all the values.
4.2.1 Definitions of UserEntered and solar panel
Table 4-1: User entered values and descriptions
Variable name
Description
UTC_Offset_UserEntered
User-entered UTC offset of site location, hours
Latitude_UserEntered
User-entered latitude of site location
Longitude_UserEntered
User-entered longitude of site location
TempCoefflscTest_UserEntered
User entered temperature coefficient of Isc of the test panel.
Default = 0.105%/°C. Configured at the factory if value is
provided
TempCoefflscRef_UserEntered
User entered temperature coefficient of Isc of the reference
panel. Default = 0.105%/°C. Configured at the factory if value
is provided
IscTeststc_UserEntered
Published panel short-circuit current (Isc) of the test panel at
STC
IscRefstc_UserEntered
Published panel short-circuit current (Isc) of the reference
panel at STC
HrSolNoonOffset_UserEntered
User editable field that defines the number of hours before
and after solar noon that will include measurements for
calculating soiling index. Default = 2 hours
GeffThreshold_UserEntered
User editable field that defines the minimum effective
irradiance for calculating soiling index. Default = 500 W/m
2
Stable_Data_Count_UserEntered
User editable field that defines the minimum number of
stable data counts for (re)calculating soiling index. Also the
minimum number of stable data counts for performing an
offset correction. Default = 20