NOTE:
This procedure (maintenance button press greater than 5 s) is only used to remove an offset
when both panels are clean. This is not the same as the regular maintenance procedure
(maintenance button press < 5 s after reference panel is cleaned).
If the maintenance button is inadvertently pressed for more than 5 s and the offset correction
is not needed, the process can be aborted by accessing the DustVue through the assigned IP
address and using
Device Configuration Utility
to reset the Offset Correction variable to 0
before midnight.
9. Maintenance
For more accurate soiling-rate results, clean the reference (clean) panel as often as the
pyranometer is cleaned, a minimum of once per week. Clean with distilled water and a lint-free
cloth. After cleaning, press the Maintenace button and hold for less than 5 s.
DustVue
14