CS451/CS456 Submersible Pressure Transducer
13
programming code to start a new program or add to an existing custom
program.
Short Cut
cannot edit programs after they are imported and edited
in
CRBasic Editor
.
A
Short Cut
tutorial is available in Section
. To import
Short
Cut
code into
CRBasic Editor
to create or add to a customized program, follow
the procedure in Appendix
Importing Short Cut Code into CRBasic Editor
(p. A-1)
. Programming basics for CRBasic data loggers are provided in the
following sections. Complete program examples for select CRBasic data
loggers can be found in Appendix
(p. B-1)
. Programming
basics and programming examples for Edlog data loggers are provided at
www.campbellsci.com\old-manuals
The
SDI12Recorder()
measurement instruction programs CRBasic data
loggers to measure the CS451/CS456 sensor.
For most CRBasic data loggers, the
SDI12Recorder()
instruction has the
following syntax:
SDI12Recorder
(Destination, SDIPort, SDIAddress, “SDICommand”,
Multiplier, Offset, FillNAN, WaitonTimeout)
For some of the SDI-12 commands, the
Destination
parameter must be an
array. The required number of values in the array depends on the command
FillNAN
and
WaitonTimeout
are optional parameters (refer to CRBasic Help
for more information).
(p. 17)
, lists the SDI-12 commands, the returned
data, and other information about the SDI-12 instruction.
7.5 Changing Nose Cones
The procedure to change or replace a nose cone follows:
1. Remove the two screws (FIGURE
2. Pull the nose cone away from the sensor housing.
3. Insert the new nose cone into the sensor housing.
4. Replace and tighten the screws.