![Campbell CS125 Instruction Manual Download Page 39](http://html1.mh-extra.com/html/campbell/cs125/cs125_instruction-manual_3803392039.webp)
User Guide
29
8
Baud rate
0-6
Baud rate for the main RS232/RS485 interface
0 = 115200 bps
1 = 57600 bps
2 = 38400 bps
3 = 19200 bps
4 = 9600 bps
5 = 2400 bps
6 = 1200 bps
2
9
Serial Number
-
Internal serial number for the CS125.
(
Read only)
-
10 Visibility Unit
M or F
Unit the visibility value will be presented as
M = metres
F = feet
M
11 Message Interval
1-3600
Interval in seconds between outputs in continuous
mode. This value has no effect if polled mode has
been selected in ‘Measurement mode’
60
12 Measurement mode
0-1
Selects polled or continuous modes.
In continuous mode the sensor will output a string
in the format as set by ‘Message Format’ at
regular intervals as defined by ‘Continuous
Interval’.
0 = Continuous mode
1 = Polled mode
0
13
Message Format
0-11
Output message
0, 1 and 2 = Basic, partial or full visibility
messages
3, 4 and 5 = Basic, partial or full SYNOP
messages
6, 7 and 8 = Basic, partial or full METAR
messages
9,10 and 11 = Basic, partial or full SYNOP
messages
5
14 Serial port protocol
0-1
Selects the physical serial interface
0 = RS232 mode
1 = RS485 mode
0
15 Averaging period
1 or 10
The period of time that the visibility value should
be averaged over. Either one minute or ten.
1
16 Sample timing
1-60
Used to define the time interval between sampling
the volume. It is recommended that this value is
left at one except when very low power demands
are needed. For example:
1 = Sample every second
2 = Sample one second in every two
3 = Sample one second in every three etc.
1
17 Dew heater override
0-1
0 = Allow the CS125 to automatically control the
dew heaters
1 = Turn the dew heaters off
0
18
Hood heater override
(1)
0-1
0 = Allow the CS125 to automatically control the
hood heaters
1 = Turn the hood heaters off
0
Summary of Contents for CS125
Page 2: ......
Page 6: ......
Page 10: ...vi...
Page 60: ......
Page 62: ...CS125 Present Weather Sensor A 2...
Page 64: ...CS125 Present Weather Sensor B 2...
Page 75: ......