![background image](http://html1.mh-extra.com/html/campbell/am16-32b/am16-32b_product-manual_3755840029.webp)
AM16/32B Relay Multiplexer 26
FIGURE B-4. Full-bridge measurement
Table B-4. Wiring for Load Cells CRBasic Example
CR1000X
AM16/32B in 4X16 mode
Sensors
Control and COM
terminals
Measurement
terminals
VX1
COM ODD H
Odd-numbered H
Excitation
⏚
(Signal Ground)
COM ODD L
Odd-numbered L
Ground
1H
COM EVEN H
Even-numbered H
High
1L
COM EVEN L
Even-numbered L
Low
⏚
COM
Shield
12V
12V
G
G
C1
CLK
C2
RES
B.5
CS616 measurement
Example
AM16-32B Example B-7 CR1000X CS616 measurements.CR1X
is a CR1000X program
measuring 48 Campbell Scientific CS616 water content reflectometers through an AM16/32B. With
minor adjustments, this program can be used with the CR6, CR3000, CR800-series, or CR1000. The
AM16/32B must be in 4x16 mode. See
AM16-32B Example B-7 CR1000X CS616 measurements.CR1X
and
for the program and wiring diagram, or use
Short Cut to create the program.