53/66
Read all instructions completely before using this product..
To avoid electric shock:
• Use caution during installation and use of this product; high voltages and currents may be
present in circuit under test.
• This product must be used only by qualified personnel practising applicable safety precautions.
• Wear protective clothing and gloves as required.
• Do not install this product on live conductors.
• Always de-energise circuit under test before installing flexible measuring head. Always
inspect the electronics unit, connecting cable, and flexible probe for damage before using this
product.
• Do not use product if damaged.
• Always connect electronics unit to display device before installing the flexible measuring head.
• Always wear protective clothing and gloves if hazardous live parts are present in the
installation where the measurement is carried out.
• If the probe is used in a manner not specified by the manufacturer the protection provided by
the equipment may be impaired.
C2 Introduction
The ACP 3000_5 and ACP 3003_5 are AC current probes utilising Rogowski principle.
They can be used to measure AC current up to 3000A, when used with oscilloscopes, recorders
or data loggers. The flexible probe allows current measurements on conductors that are hard to
reach.
The probes provide a 3V full scale AC output proportional to the current being measured with
three selectable ranges.
Fig. 1
1.
Probe Coupling
2.
Probe Output Cable
3.
Flexible Current Probe
4.
Power On / Range Switch
5.
RED LED – Overload
6.
Integrator Housing
7.
Output Cable