4.12 Temperature calibration function
Click Temperature Calibration on the
function setting screen to enter the
temperature calibration interface.
If the furnace temperature deviates
from the set temperature, use the
temperature calibration function to
ensure the temperature accuracy.
According to Step 1, click Temperature
Test, the tray will drop to the lowest
end. Place the temperature measuring
ring (PTCR-UTH) in the middle of the
firing platform, and click Confirm. After
the furnace temperature reaches 800℃
and retain it for 60 minutes. Then the
tray gradually drops to the lowest end.
After the temperature measuring ring
drops to room temperature, measure
its outer diameter with the caliper for
three times and take the average value.
You can find the actual temperature
measured by the temperature
measuring ring according to the
temperature table of the temperature
measuring ring.