7
LIST OF MODBUS REGISTERS
The list below includes all available addresses on the PyroMiniBus sensor.
R = Read, W = Write
Address
Length (words)
Description
R/W
0x00
1
Sensor Type (17 for PyroMiniBus)
R
0x01
1
Field of view 0 for 2:1; 1 for 20:1
R
0x02
2
Serial number
R
0x04
1
Modbus slave ID (slave address)
R/W
0x05
1
Reflected Energy Compensation 0 for Off; 1 for On
R/W
0x06
1
Reflected Temperature
R/W
0x07
1
Emissivity Setting (1 LSB = 0.0001)
Minimum 0.2000, Maximum 1.0000
R/W
0x08
1
Hold Mode 0 for Off; 1 for Peak; 2 for Valley
R/W
0x09
1
Hold Period (1 LSB = 0.1 seconds)
Minimum 0.1 seconds, Maximum 1200.0 seconds
R/W
0x0A
1
Average Period (1 LSB = 0.1 seconds)
Minimum 0.1 seconds, Maximum 60.0 seconds
R/W
0x0B
1
Average Temperature
R
0x0C
1
Minimum Temperature
R
0x0D
1
Maximum Temperature
R
0x0E
1
Filtered Temperature
R
0x0F
1
Unfiltered Temperature
R
0x10
1
Sensor Temperature
R
0x11
1
Status (bits active high)
Bit 0: Measurement error
Bit 1: Sensor temperature low
Bit 2: Sensor temperature high
Bit 3: Object temperature low
Bit 4: Object temperature high
R
0x16
1
Reflected Temperature (not saved to non-volatile
memory)
R/W
0x17
1
Emissivity Setting (not saved to non-volatile
memory)
R/W
Notes:
1 All temperature are in tenths of degrees C
2 With the exception of addresses 0x16 and 0x17, all write operations are saved to non-
volatile memory
3 For further information please refer to http://www.modbus.org/specs.php
4 Use address 255 to communicate with any connected sensor (only one sensor
connected)
5 Use address 0 to broadcast to all connected sensors (no response expected)