26
Questions about Cala Trio? Visit CalaTrio.com/HCPFAQs
LBL-5122 Rev C NOV 2019
Cala Trio Stimulator claims compliance for
electromagnetic compatibility in conjunction with Cala
Trio Base Station.
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received,
including interference that may cause undesired operation.
Do not tamper with, modify, or attempt to perform
maintenance or servicing on Cala Trio.
RF Exposure Evaluation Guidance for Mobile Conditions
A minimum separation distance of 20 cm is required
between the Cala Trio Base Station and nearby person to
qualify for mobile exposure limits.
GENERAL RADIO COMPLIANCE
ITEM
SPECIFICATION
Stimulator
Bluetooth
FCC ID: 2AA9B05
Radio Frequency
(RF) Range
2402 MHz to 2480 MHz
Base Station
FCC ID: 2AT2D-BASE0017
Radio Frequency
(RF) Range
13.56 MHz
2402 MHz to 2480 MHz
LTE Module
FCC ID: XPY2AGQN4NNN
Radio Frequency
(RF) Range
Device operates within approved frequencies
overlapping with the following cellular bands:
LTE 2,1900 PCS UP | LTE 25,1900+ UP | LTE
35,TD PCS Lower DOWN | UMTS CH 2 UP |
UMTS CH 25 UP | UMTS CH 35 DOWN |
GUIDANCE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC EMISSIONS
Cala Trio is intended for use in the electromagnetic environment specified below. The customer or the user of Cala Trio should assure that it is
used in such an environment.
EMISSIONS TESTS
COMPLIANCE
RF emissions
CISPR 11
Group 1
Cala Trio uses RF energy only for its internal function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
Cala Trio is suitable for use in all establishments, including domestic establishments and those
directly connected to the public low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/
Flicker emissions
Complies
GUIDANCE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC IMMUNITY
Cala Trio is intended for use in the electromagnetic environment specified below. The customer or the user of Cala Trio should assure that it is
used in such an environment.
IMMUNITY TEST
IEC 60601 TEST LEVEL
COMPLIANCE LEVEL
ELECTROMAGNETIC ENVIRONMENT GUIDANCE
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV for power supply
lines
Mains power quality should be that of a typical commercial
or hospital environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality should be that of a typical commercial
or hospital environment.
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.
ELECTROMAGNETIC COMPATIBILITY
DECLARATION
11.0