background image

PN MN234 C | User Manual    3 - ENG

Liberator

Specifications

•  Mode of Operation: Continuous Flow

•  Type of Protection Against Electrical Shock: Internally Powered Equipment

• 

 Degree of Protection Against Electrical Shock: Type BF Applied Part

•  IP21 Classification According to the Degree of Protection Against Ingress of Water: Internal protec-

tion against ingress of solid foreign objects greater than or equal to 12.5 mm in diameter and ingress 

of vertically dripping water.

• 

 Equipment not suitable for use in the presence of flammable mixtures

Product Specifications

Liberator 20

Liberator 30

Liberator 37

Liberator 45

Liberator 60

LOX Capacity

21.0 L 
50.7 lb  
(23 kg)

31.0 L 
74.8 lb  
(33,9 kg)

37.3 L 
91.0 lb  
(41,3 kg)

45.7 L 
110.3 lb  
(50,04 kg)

60.2 L 
145.3 lb  
(65,9 kg)

Gaseous  
Equivalent  
Capacity

17,337 L

25,580 L

31,121 L

37,724 L

49,679 L

Weight, Empty

39 lb  
(17,96 kg)

47.5 lb  
(21,54 kg)

50 lb  
(22,68 kg)

55 lb  
(24,95 kg)

75.4 lb  
(34,19 kg)

Weight, Filled

89.7 lb  
(40,69 kg)

122.77 lb  
(55,68 kg)

141.01 lb  
(63.98 kg)

165.32 lb  
(74.99 kg)

220.68 lb  
(100,1 kg)

Height

24.5 in.  
(622 mm)

29.5 in.  
(750 mm)

32.75 in.  
(832 mm)

37 in. (940 mm)

39 in. (990 mm)

Diameter

14 in. (356 mm) 14 in. (356 mm) 14 in. (356 mm) 14 in. (356 mm)

16 in. (406 mm)

Typical use time 
at 2 LPM

6 days 12 hrs

9 days 9 hrs

11 days 14 hrs

14 days 2 hrs

18 days 2 hrs

Operating  
Pressure

20 psi (137 kPa) 20 psi (137 kPa) 20 psi (137 kPa) 20 psi (137 kPa)

20 psi (137 kPa)

Normal  
Evaporation Rate

1.6 lb/ day   
(0,73 kg/day)

1.6 lb/ day   
(0,73 kg/day)

1.6 lb/ day   
(0,73 kg/day)

1.6 lb/ day   
(0,73 kg/day)

1.65 lb/ day  
(0,75 kg/day)

Standard Flow 
Control Range

Off, .25, .5, .75, 
1, 1.5, 2, 2.5, 3, 
4, 5, 6, 8, 10, 12, 
15 LPM

Off, .25, .5, .75, 
1, 1.5, 2, 2.5, 3, 
4, 5, 6, 8, 10, 12, 
15 LPM

Off, .25, .5, .75, 
1, 1.5, 2, 2.5, 3, 
4, 5, 6, 8, 10, 12, 
15 LPM

Off, .25, .5, .75, 
1, 1.5, 2, 2.5, 3, 
4, 5, 6, 8, 10, 12, 
15 LPM

Off, .25, .5, .75, 
1, 1.5, 2, 2.5, 3, 
4, 5, 6, 8, 10, 12, 
15 LPM

Flow Rate  
Accuracy*

+/- 10%

+/- 10%

+/- 10%

+/- 10%

+/- 10%

* This accuracy is only at 70F and 14.7 psig and with a calibrated accurate mass flow meter.

Summary of Contents for Liberator 30

Page 1: ...Liberator User Manual English...

Page 2: ...tive 93 42 EEC concerning medical devices Authorized representative in the European Community This device complies with the requirements of Directive 93 42 EEC concerning medical devices It bears the...

Page 3: ...kg 47 5 lb 21 54 kg 50 lb 22 68 kg 55 lb 24 95 kg 75 4 lb 34 19 kg Weight Filled 89 7 lb 40 69 kg 122 77 lb 55 68 kg 141 01 lb 63 98 kg 165 32 lb 74 99 kg 220 68 lb 100 1 kg Height 24 5 in 622 mm 29 5...

Page 4: ...bite Liquid and gaseous oxygen though nonflammable cause other materials to burn faster than normal This hazard along with the low temperature of liquid oxygen warrants certain safety precau tions Kee...

Page 5: ...tor is offered as a top fill model or a dual top fill and side fill unit They are filled by your health care provider The portable provides an am bulatory source of oxygen for an extended period of ti...

Page 6: ...DOES NOT SEPARATE EASILY DO NOT USE FORCE THE UNITS MAY BE FROZEN TOGETHER LEAVE THE UNITS CONNECTED AND WAIT UNTIL THEY WARM UP THEN THEY WILL SEPARATE EASILY DO NOT TOUCH ANY FROSTED PARTS Caution S...

Page 7: ...Indicator lights up when the button is depressed inform your health care pro vider the next time your Liberator is filled Gen 4 Meter 5 Install the DISS extension 6 Either a Attach cannula to the DIS...

Page 8: ...utton to display level If level is displayed and Low Battery Indicator is not il luminated battery level is acceptable If the Low Battery Indicator lights up when the contents button is depressed call...

Page 9: ...to time to relieve excess pressure especially soon after filling If hissing is persistent or abnormal this could indicate excess pressure being vented off or a leak in the system Contact service prov...

Page 10: ...local city or town offices for instructions on proper disposal of the battery WEEE and RoHS This symbol is to remind the equipment owners to return it to a recycling facility at the end of its life pe...

Page 11: ...tic establishments and those directly connected to the Voltage fluctuations public low voltage power supply network that supplies flicker emissions Not applicable buildings used for domestic purposes...

Page 12: ...d device IEC 610004 4 1 kV for Not applicable input output linesNo data input output lines 1 kV line s Surge to line s Not Applicable Not Applicable IEC 61000 4 5 2 kV line s DC powered device to eart...

Page 13: ...rya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequ...

Page 14: ...14 ENG PN MN234 C User Manual Liberator Notes...

Page 15: ...PN MN234 C User Manual 15 ENG Liberator Notes...

Page 16: ...and with no further obligation or consequence All rights not expressly stated herein are reserved by us as applicable www cairemedical com CAIRE Inc 2200 Airport Industrial Dr Ste 500 Ball Ground GA...

Reviews: