PN MN232-C4 J | User Manual ENG - 31
FreeStyle Comfort
Guidance and Manufacturer’s Declaration–Electromagnetic Immunity
The FreeStyle Comfort is intended for use in the electromagnetic environment specified below. The customer or
the user of the FreeStyle Comfort should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment –
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
10 Vrms
150 kHz to 80 MHz
10 V/m
80 MHz to 2700 MHz
10 Vrms
10 V/m
Portable and mobile RF
communications equipment should
be used no closer to any part of the
FreeStyle Comfort, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
P
d
2
.
1
P
d
2
.
1
= 80 MHz to 800 MHz
P
d
2
.
1
= 800 MHz to 2700 MHz
where
P
is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and
d
is the
recommended separation distance in
metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should be
less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic
site survey should be considered. If the measured field strength in the location in which the FreeStyle Comfort
is used exceeds the applicable RF compliance level above, the FreeStyle Comfort should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as
re-orienting or relocating the FreeStyle Comfort.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Summary of Contents for FreeStyle Comfort
Page 346: ...FreeStyle Comfort 346 GRE PN MN232 C4 J AC DC autoSAT...
Page 347: ...FreeStyle Comfort PN MN232 C4 J GRE 347 FreeStyleComfort FREESTYLE COMFORT 1 2...
Page 348: ...FreeStyle Comfort 348 GRE PN MN232 C4 J DOT UN AC GFI...
Page 349: ...FreeStyle Comfort PN MN232 C4 J GRE 349 3 10 WWW P65WARNINGS CA GOV AC AC...
Page 350: ...FreeStyle Comfort 350 GRE PN MN232 C4 J DC DC RF 25 50 20 C 20 C 4 F 68 F AC DC...
Page 359: ...FreeStyle Comfort PN MN232 C4 J GRE 359 AC DC 3 6 3...
Page 360: ...FreeStyle Comfort 360 GRE PN MN232 C4 J 1 1 20 2 2 1 20 2 3 1 20 2 4 1 30 15 5 1 30 6...
Page 361: ...FreeStyle Comfort PN MN232 C4 J GRE 361 7 1 30 5 8 OCD 1 30 9 1 10 1 60 40bpm 11 12 13 1...
Page 362: ...FreeStyle Comfort 362 GRE PN MN232 C4 J autoSAT DC autoSAT DC 1 CAIRE 2 5 1 CAIRE 2 1...
Page 363: ...FreeStyle Comfort PN MN232 C4 J GRE 363 5 1 1...
Page 367: ...FreeStyle Comfort PN MN232 C4 J GRE 367 FreeStyle Comfort AC DC AC DC AC DC...
Page 498: ...498 ENG PN MN232 C4 J User Manual FreeStyle Comfort Notes...
Page 499: ...PN MN232 C4 J User Manual ENG 499 FreeStyle Comfort Notes...