PN MN232-C4 J | User Manual ENG - 19
FreeStyle Comfort
Alarm Conditions
Priority # Alarm
Description
Display
Audible
Signal
More Details / Suggested Action
1
System
Malfunction
Battery High
Temperature
1 beep
every
20 sec
Unit will alarm when battery tempera-
ture is approaching safety shutdown
threshold.
If shutdown threshold is reached, unit
will shut down. If external power is
connected, alarm screen will display and
audible alarm will continue for 2 min.
Test with battery at room temperature. If
problem persists, contact provider.
2
System
Malfunction
High
Temperature
1 beep
every
20 sec
Unit will alarm when internal tempera-
ture is approaching shutdown threshold.
If shutdown threshold is reached, unit
will shut down and alarm screen will
display and audible alarm will continue
for 2 min.
Ensure there are no blockages on any
vents of the unit. If problem persists,
contact provider.
3
System
Malfunction
High
Pressure
1 beep
every
20 sec
If shutdown threshold is reached, unit
will shut down and alarm screen will
display and audible alarm will continue
for 2 min.
If problem persists, contact provider.
4
No Breath
Detected
1 beep
every
30 sec
No breath detected for 15 seconds.
Activated after first breath or warm up
screen is off, whichever comes sooner.
Check cannula for kinks or leaks. Increase
the sensitivity setting. If problem persists,
contact provider.
5
Low Battery
Shut Down
1 beep
every
30 sec
Alarms until battery runs out of power
and unit shuts down.
Charge battery.
6
Warm Up
none
Unit will start by auto-dosing until a
breath is detected or Warm up ends,
whichever is sooner.
Summary of Contents for FreeStyle Comfort
Page 346: ...FreeStyle Comfort 346 GRE PN MN232 C4 J AC DC autoSAT...
Page 347: ...FreeStyle Comfort PN MN232 C4 J GRE 347 FreeStyleComfort FREESTYLE COMFORT 1 2...
Page 348: ...FreeStyle Comfort 348 GRE PN MN232 C4 J DOT UN AC GFI...
Page 349: ...FreeStyle Comfort PN MN232 C4 J GRE 349 3 10 WWW P65WARNINGS CA GOV AC AC...
Page 350: ...FreeStyle Comfort 350 GRE PN MN232 C4 J DC DC RF 25 50 20 C 20 C 4 F 68 F AC DC...
Page 359: ...FreeStyle Comfort PN MN232 C4 J GRE 359 AC DC 3 6 3...
Page 360: ...FreeStyle Comfort 360 GRE PN MN232 C4 J 1 1 20 2 2 1 20 2 3 1 20 2 4 1 30 15 5 1 30 6...
Page 361: ...FreeStyle Comfort PN MN232 C4 J GRE 361 7 1 30 5 8 OCD 1 30 9 1 10 1 60 40bpm 11 12 13 1...
Page 362: ...FreeStyle Comfort 362 GRE PN MN232 C4 J autoSAT DC autoSAT DC 1 CAIRE 2 5 1 CAIRE 2 1...