PN 13402391 D | User Manual    11 - ENG

EMS 20

Safety

WARNING: PORTABLE RF COMMUNICATIONS EQUIPMENT (INCLUDING PERIPHERALS SUCH AS 
ANTENNA CABLES AND EXTERNAL ANTENNAS) SHOULD BE USED NO CLOSER THAN 30 CM (12 
INCHES) TO ANY PART OF THE EMS 20, INCLUDING CABLES SPECIFIED BY THE MANUFACTURER. 
OTHERWISE, DEGRADATION OF THE PERFORMANCE OF THIS EQUIPMENT COULD RESULT.

WARNING: USE OF ACCESSORIES, TRANSDUCERS AND CABLES OTHER THAN THOSE SPECIFIED OR 
PROVIDED BY THE MANUFACTURER OF THIS EQUIPMENT COULD RESULT IN INCREASED ELECTROMAG-
NETIC EMISSIONS OR DECREASED ELECTROMAGNETIC IMMUNITY OF THIS EQUIPMENT AND RESULT IN 
IMPROPER OPERATION.
WARNING: USE OF THIS EQUIPMENT ADJACENT TO OR STACKED WITH OTHER EQUIPMENT SHOULD 
BE AVOIDED BECAUSE IT COULD RESULT IN IMPROPER OPERATION. IF SUCH USE IS NECESSARY, THIS 
EQUIPMENT AND THE OTHER EQUIPMENT SHOULD BE OBSERVED TO VERIFY THAT THEY ARE OPERATING 
NORMALLY.

Caution: Medical Electrical Equipment needs special precautions regarding Electromagnetic 
compatibility (EMC) and needs to be installed and put into service according to the EMC infor-
mation provided in this manual.
Caution: Portable and mobile radio frequency (RF) communications equipment can affect 
Medical Electrical Equipment.
Caution: The EMS 20 should not be used adjacent to or stacked with other equipment and that 
if adjacent or stacked use is necessary, the EMS 20 should be observed to verify normal opera-
tion in the configuration in which it will be used.

Table 1

Guidance and Manufacturer’s declaration—electromagnetic emissions

The EMS 20 is intended for use in the electromagnetic environment specified below. The customer or the 

user of the EMS 20 should assure that it is used in such an environment.

  Emissions test 

Compliance  Electromagnetic environment—guidance

  RF emissions 

Group 1 

The EMS 20 uses RF energy only for internal function.

  CISPR 11 

 

Therefore, its RF emissions are very low and are not likely to 

   

 

cause any interference in nearby electronic equipment.

  RF emissions CISPR 11  Class B 

  

  Harmonic emissions  

 

The EMS 20 is suitable for use in all establishments, including

  IEC 61000-3-2 

Not applicable  domestic establishments and those directly connected to the 

  Voltage fluctuations/ 

 

public low-voltage power supply network that supplies

   flicker emissions  

Not applicable  buildings used for domestic purposes.

  IEC 61000-3-3

Summary of Contents for EMS 20

Page 1: ...EMS 20 User Manual English...

Page 2: ...ual must be read Reg M002 Keep away from open flame fire sparks Open ignition source and smoking prohibited Reg P003 Do not smoke near unit or while operating unit Reg P002 Warning Reg W001 ADR Europe...

Page 3: ...y 21 0 L 50 7 lb 23 kg Gaseous Equivalent Capacity 17 337 L Weight Empty 39 lb 17 96 kg Weight Filled 89 7 lb 40 69 kg Height 24 5 in 622 mm Diameter 14 in 356 mm Typical use time at 2 LPM 6 days 12 h...

Page 4: ...The unit contains liquid oxygen which is extremely cold almost 300 F 184 C Exposure to such a low temperature can cause severe frostbite Liquid and gaseous oxygen though nonflammable cause other mater...

Page 5: ...tion OI contains the instruc tions for using the EMS system as well as informa tion regarding available accessories For detailed instructions regarding accessories please see their respective document...

Page 6: ...ne by pressing QDV poppet at end of transfer line against non painted surface for a few seconds 4 Connect pressure gauge to oxygen outlet and open outlet valve 5 Open vent valve on EMS unit 6 Connect...

Page 7: ...h continuous use the EMS 20 may develop an excessive amount of ice on the warming and breathing coils within the shroud You should defrost the unit between liquid oxygen fills to prevent this ice buil...

Page 8: ...between each fill to prevent freezing and possible equipment failure There are no user serviceable parts in the EMS 20s Your service provider is responsible for any mainte nance that my be required p...

Page 9: ...me to relieve excess pressure especially soon after filling If hissing is persistent or abnormal this could indicate excess pressure being vented off or a leak in the system Contact service provider f...

Page 10: ...e manual Disposal Always return EMS 20 including all components to your homecare provider for proper disposal You can also contact your local city or town offices for instructions on proper disposal o...

Page 11: ...C infor mation provided in this manual Caution Portable and mobile radio frequency RF communications equipment can affect Medical Electrical Equipment Caution The EMS 20 should not be used adjacent to...

Page 12: ...equipment which has been tested to specific test levels and over specific frequency ranges and been found compliant with regulations Table 3 Guidance and manufacturers declaration electromagnetic immu...

Page 13: ...by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of equip ment marked with the following symbol Test frequen...

Page 14: ...14 ENG PN 13402391 D User Manual EMS 20 Notes...

Page 15: ...PN 13402391 D User Manual 15 ENG EMS 20 Notes...

Page 16: ...ions specifications and or processes to its products at any time without prior notice and with no further obligation or consequence All rights not expressly stated herein are reserved by us as applica...

Reviews: