![CAI 700M Operator'S Manual Download Page 146](http://html1.mh-extra.com/html/cai/700m/700m_operators-manual_3331460146.webp)
700M HFID
6-2-15
146
All opto-isolated relays are rated for 48VDC, 0.5 Amp maximum. The user is required to limit
the drive current supplied to each input. All inputs are DC only and will not operate on AC
current.
CAUTION:
Do not connect these pins directly to both sides of a voltage power supply
as unrestricted current will damage the relay.
Example of digital output driver.
Digital Input
The analyzer’s Digital Inputs are internally pulled up to 5VDC and are operated by user
equipment connecting an input to the Digital Ground (Main, Pin 6). Note that some lines require
only momentary operation, and some selection lines are intended to be held continuously.
DRV
The calibration gas valve drivers are application-specific and intended for solenoid valve drive.
24VDC valves with a maximum wattage of 12 Watts should be used. The 24VDC used to
operate the valves should be connected to the coils and the analyzer inputs when operating will
pull the valve current to Digital Ground.
24VDC
The 24DVC output is intended for use with properly rated solenoid valves.
CAUTION:
Use of
this output for other purposes can damage the analyzer.
Analog Input
The Analog Input is reserved for factory signals.
CAUTION:
Do not connect to this input or
damage may occur.
Summary of Contents for 700M