7.3
Troubleshooting
In case of upgrading failure (e.g. STD FLASH page is corrupted), the user can try to reboot the
board: after a power cycle, the system will try to program the board automatically from the
alternative FLASH page (e.g. BKP FLASH page) if this is not corrupted as well. The user can
perform a further upgrade attempt on the STD page to restore the firmware copy.
7.3.1
PCB revision 0
NOTE THAT
old versions of the digitizer motherboard (bit[7:0] = 0x00 at register address 0xF050)
have a slightly different FLASH management. At power-on, the microcontroller loads exactly the
firmware copy from the STD page of the FLASH.
In this case, when a failure occurs during the upgrade of the STD page of the FLASH, which
compromises the communication with the DT5740, the user can perform the following
recovering procedure as first attempt:
- Force the board to reboot loading the copy of the firmware stored on the BKP page of the
FLASH. For this purpose, make sure to connect by USB link and use the Reboot function in
CAENUpgrader software by checking “Bkp” option
- Use CAENUpgrader to read the firmware revision (in this case the one of the BKP copy). If this
succeeds, it is possible now to communicate again with the board.
- Use CAENUpgrader to load again the firmware on the STD page, then power-cycle in order the
board to get operative again.
If neither of the procedures described at § 7.3 and § 7.3.1 succeeds, it is recommended to send
the board back to CAEN in repair (see § 8).