background image

 

 

 

 

 

 

 

 

 

 

Document type: 

Title : 

Re vision date: 

Re vision: 

User's Manual (MUT) 

Mod. A1735-A1835-A1735B-A1835B HV Boards 

24/03/2014 

 
 
 

NPO : 

File name: 

Number of pages: 

Page : 

00101/00:17XXx.MUTx/08 

A1735B-A1835B_REV8.DOCX 

14 

 

1.  Mod. A 1735 - A 1835  HV Boards  

1.1 

Functional description 

The Mod. A 1735 –  A 1835 are single width boards housing 12  HV channels with either 
positive (A 1735P – A 1835P)  or negative  (A 1735N  –  A 1835N)  polarity.  

28 channel double width versions (Mod. A 1735B –  A 1835B) have  either positive 
(A 1735BP  – A 1835BP)  or negative  polarity (A 1735BN  – A 1835BN)

1

The boards are  compatible with the CAEN Universal Multichannel Power Supply System 
(SYx527) 

The Mod. A 1735 -  A 1735B output channels have 1.5 kV / 7 mA full scale range (dip-
switch selectable); the Mod. A 1835  -  A 1835B output channels offer either 1.5 kV / 7 mA 
or 1.5 kV / 200 µA full scale range (dip-switch selectable). 

 If the output voltage differs from the programmed value by more than 3% of voltage full 
scale range, the channel is signalled to be either in OVERVOLTAGE or 
UNDERVOLTAGE condition. Moreover, for each channel, a voltage protection limit 
SVMAX can be fixed via software with 1 V resolution and the output voltage can not be 
programmed beyond this value.  The HV RAMP-UP and RAMP-DOWN rates may be 
selected independently for each channel in the range 1

÷

  500 V/s in 1 V/s steps.  The 

output current is monitored with 20 nA / 500 nA resolution depending on current range; if 
a channel tries to draw a current larger than its programmed limit it is signalled to be in  
OVERCURRENT condition; the SY 1527 system detects this state as a fault and reacts 
according  to the setting of the TRIP  parameter

2

, namely: 

TRIP=infinite  ( = 1000 s) 

When the set output current value is reached the channel behaves like a constant 
current generator. 

2) TRIP=finite  (< 1000 s) 

The output current keeps the set value only for programmed time interval and then is 
switched off. 

The TRIP time (i.e. the maximum time an OVERCURRENT condition is allowed to last) 
can be programmed  in 0.1 s steps. 

The maximum output voltage (VMAX Hardware) can be fixed, through a potentiometer 
located on the front panel, at the same common value for all the board channels and this 
value  can be read  out via  software. 

The boards host also a temperature sensor located on the PCB near the HV channels: 
the temperature values measured by this sensor are used to signal Over Temperature 
condition on the SYX527.   

The boards are provided with an "HV EN" input that disables the channels when it is not 
connected to ground. 

                                                                 

1

 A 1735B – A 1835B are currently out of stock and no longer produced 

2

 Refer to the SYx527 User’s Manuals for details about the TRIP Handling. 

Summary of Contents for A 1735

Page 1: ...Technical Information Manual MOD A 1735 A 1835 A 1735B A 1835B 12 28 CHANNEL HV BOARDS MANUAL REV 8 24 March 2014 Revision n 8 NPO 00101 00 17XXx MUTx 08...

Page 2: ...ight to change partially or entirely the contents of this Manual at any time and without giving any notice Disposal of the Product The product must never be dumped in the Municipal Waste Please check...

Page 3: ...ON 5 1 2 CHANNEL CHARACTERISTICS TABLE 6 1 3 FRONT PANEL 7 1 4 TECHNICAL SPECIFICATIONS 8 1 4 1 Packaging 8 1 4 2 External connections 8 1 4 3 Displays 9 1 4 4 Other components 9 2 SAFETY INFORMATION...

Page 4: ...5B A1835B_REV8 DOCX 14 4 LIST OF FIGURES FIG 1 1 MOD A 1735 A 1835 LEFT AND MOD A 1735B A 1835B RIGHT FRONT PANEL 7 FIG 1 2 MOD A 1735B A 1835B HVMULTIPIN CONNECTOR 9 FIG 3 1 A MOD A1735 A 1835 SIDE V...

Page 5: ...DOWN rates may be selected independently for each channel in the range 1 500 V s in 1 V s steps The output current is monitored with 20 nA 500 nA resolution depending on current range if a channel tri...

Page 6: ...nA depending on current range Mod A 1835 A 1835B VMAX hardware 0 1 5 kV common for all the board channels VMAX hardware accuracy 2 of FSR VMAX software 0 1 5 kV settable for each channel VMAX softwar...

Page 7: ...T P U T S SCALER POSITIVE 12 CH Mod V560E Mod A 1835P HV EN input connector Channel output connector Channel LED SCALER Mod V560E Mod A1835BP 28 CH HV EN MAXV GND GND GND GND GND NC NC NC NC NC 6 NC...

Page 8: ...35 A 1835 Mechanical specifications HV coaxial connectors Radiall SHVR317580 type Electrical specifications high voltage output according to specifications given in Table 2 1 p 6 CH 0 27 A 1735B A 183...

Page 9: ...positive polarity version yellow LEDs for negative polarity version HV ON LED A 1735B A1835B Function it lights up if at least one output channel is on Type red LED for positive polarity version yell...

Page 10: ...ck or fire hazard do not apply a voltage to a load that is outside the range specified for that load Avoid Electric Shock To avoid injury or loss of life do not connect or disconnect cables while they...

Page 11: ...product The following symbols may appear on the product DANGER ATTENTION High Voltage Refer to Manual 2 3 Installation The Mod A 1735 A 1835 are single width boards which can be inserted in any slot...

Page 12: ...tput channel it is possible through the SYX527 system to perform the following operations Assign to channel a symbolic name Set output voltage VSET Set max output current ISET Set output voltage softw...

Page 13: ...o enable the board s operation Provide the HV EN connector with a 50 Ohm termination Provide short circuit between Pin 51 and Pin 52 IINT and IEXT of the HV Multipin Connector see Fig 2 2 this is a sa...

Page 14: ...els Dip Switch side view T A B C 4 3 2 1 A B Dip Switch T A B C 4 3 2 1 Channels Channels Channels Channels Channels Channels Channels Fig 3 1 A Mod A1735 A 1835 side view B Mod A1735B A 1835B top vie...

Reviews: