Cadex C5100B Battery Tester User Guide
19
2016 Cadex Electronics Inc.
Possible Fault Messages
Contact Cadex Technical Support at (Canada) 1-604-231-7777, (USA Toll Free) 800-565-5228,
or (Europe) +49 (0) 6105 975 11-0, or [email protected] for further information.
Fault Codes
Code Name
Description
-3115 Target Capacity Not Met
Battery Capacity < Target Capacity
-3120 Over Voltage
Battery voltage too high
-3121 Battery Reversed
Battery inserted backwards
-3122 Battery Shorted
Battery has an internal Short.
-3129 Intermittent Battery
Battery processing interrupted more than 5 times.
-3142 Discharge Timeout
Discharge longer than specified timeframe.
-3144 Charge Timeout
Charge longer than specified timeframe
-3150 Thermistor Failure
Unable to read temperature
-3159 Hot Battery
Battery is too hot to start processing.
-3160 Bad Fuse Or Driver
No current for more than 30 seconds.
-3201 Cold Battery
Battery is too cold to start processing.
-3202 Battery Low
Battery voltage is too low to charge safely.
Warning Codes
Code Name
Description
-2159 Hot Battery
Battery became too hot during processing. Processing is put on
hold and will continue when the battery cools down.
-2201 Cold Battery
Battery became too cold during processing. Processing is put on
hold and will continue when the battery warms up.
-2203 System Over Temperature
The system became too hot during processing. Process is put on
hold and will continue when the system cools down.
-2206 System Under Temperature
The system became too cold during processing. Processing is put
on hold and will continue when the system warms up.