PREPARATION AND USE
When first used o
r always after long time of non-use, we recommend disinfection of the nebulisation
chamber, the mask and mouthpiece, so to avoid propagation and infection from miscellaneous bacteria. Then
follow the instructions in the Cleaning and Disinfection paragraph of this manual and proceed with the
operations described below.
DO NOT INSERT OR NEBULISE substances other than the drugs or medical devices intended for
aerosol nebulisation in this device. DO NOT USE WITH Medical Device and/or with dense solutions
containing pure hyaluronic acid or dissolved in a high percentage, essential oils, fragrances, etc., or
sediment loads (e.g. Pure Thermal Waters) that can block or permanently damage the device.
Use with pure hypertonic solutions (e.g. Seawater) can block the device if not properly cleaned
immediately afterwards each use.
1
- DRUG INSERTION
FIG A
FIG B
FIG C
A) Open the drug chamber cover by releasing the appropriate safety lever with your fingers.
B) Insert the prescribed drug(s) in compliance with the instructions of your doctor and the drug manufacturer.
ATTENTION:
The drug chamber is equipped with a graduated scale indicating the maximum volume of drug that
can be inserted for each individual application. Th
is volume is 8ml (8cc). Even if the maximum capacity of the
chamber allows up to 12 ml of drug, it is recommended not to exceed the maximum value of 8 ml indicated on the
graduated scale.
C) Close the drug chamber cover again, making sure to have properly engaged the safety lever.
Do not drop medication on the Main Unit. If you drop medication, immediately wipe it off with gauze.
If you use the unit while it is wet, it may cause trouble or you may suffer electric shock.
On the stop perimeter of the cha
mber cover, a circular transparent gasket is fixed in its appropriate seat and
must never be removed, not even during cleaning operations. Always check that it is in place and in case of
accidental loss, do not use the product in order to prevent drug spillage and leakage. If the gasket is lost, you
must replace the nebulisation chamber.
Summary of Contents for KIWI Plus
Page 1: ...1...
Page 6: ...6 KIWI Plus MESH PC 200 0 1 ml 93 MMAD 1 51 m CA MI 1 2 3...
Page 7: ...7 4 5 6 KIWI Plus 7 8 9 ISO 10993 1 10 EN 60601 1 11 12 13 14...
Page 8: ...8 15 93 42 EEC...
Page 9: ...9 1 8 ml 8 cc 12 ml 8 ml...
Page 10: ...10 2 4 L S...
Page 12: ...12 3 2 100 220 28 1 1 2 5 ml NaCl 0 9 300 1 100 3 2 1 DC 4 100 1 300...
Page 13: ...13 LED 4 20 90...
Page 14: ...14 1 1 1...
Page 15: ...15 2 1 2 5 3 30 4 Y 10...
Page 17: ...17 2 180 5 2 1 0 3 2 3 4 5 1 2 3 4 5...
Page 20: ...20 CA MI Srl 24 CA MI srl CA MI srl 93 42 BF II AC DC...
Page 21: ...21 IPX2 1 2 15 2006 66 2012 19...
Page 57: ...57...