background image

EPS-300 

  7 

run, a ~3 second beep sounds and the display changes to “OFF”. To reset the power supply 

press any button except the RUN button. 

PFD:

 The Power Failure Detection feature will save the time, voltage and current settings in 

the event of a power outage and automatically continue the run when the power is restored. 

To alert the user that a power failure occurred during the run, the error message “E6” flashes 

at the end of the run instead of the usual “OFF” message. The “E6” error message means 

that  despite  the  power  outage,  the  power  supply  successfully  supplied  the  programmed 

voltage and current for the entire length of time that was programmed. The time shown at 

the end of the run is the time the gel ran since power was restored, not the total time of the 

run. The default settings for Power Failure Detection are: 

   

 

Run   

 

 

Time Setting   

PFD   

Code 

   

 

Timed Run 

 

1 – 1440 minutes 

ON   

PF1 

   

 

Continuous run   

 

 

 

OFF   

PF0   

 

The Power Failure Detection feature can be deactivated during a timed run. 

  Press the MODE button to move the indicator light under clock symbol. 

  Press the UP and DOWN buttons to enter in the run time in minutes. 

  Press both the UP and DOWN buttons at the same time. The display will briefly 

show “PF0”, indicating that the PFD mode is disabled. 

  Press both the UP and DOWN buttons at the same time again. The display will 

show “PF1”, indicating that the PFD mode is enabled. 

 

8.  Start the run: 

Press the RUN button. The green LED light above the RUN button indicates 

that voltage and current are being applied to the output jacks. The green LED light above the 

CONST button indicates which parameter (V or mA) is limiting. The LED display shows the 

numerical value of the actual V or mA output, or the time (Press the MODE button to toggle 

between the values). The timer will count up in the case of a continuous run and count down 

if the timer was set for a specific timed duration. 

 

9.  Stop the run:

 A timed run will automatically stop with a warning beep as described above. 

In the case of a continuous run, press RUN button to terminate the run (The green LED light 

above the RUN button will turn off indicating that the voltage and current which was being 

applied  to  the  output  jacks  were  stopped.).  The  display  reads  “OFF”.  To  reset  the  power 

supply press any button except the RUN button. 

Summary of Contents for EPS-300 II

Page 1: ...EPS 300 II and EPS 300 IIV Power Supply Installation and Operation Manual Version 1 0 This instrument is intended for laboratory use only...

Page 2: ...ct Description B 2 1 EPS 300 Power Supply Hardware Overview C Installation of EPS 300 Power Supply 5 C 1 Package List C 2 Setup and Operation of the EPS 300 Power Supply C 2 1 Unpacking C 2 2 Setup an...

Page 3: ...anty EPS 300 is warranted to be free from defects in materials or workmanship for a period of one year from the original invoice date under normal usage Any defects occurring during warranty period C...

Page 4: ...t the power supply using the AC power cord provided with the power supply or a similar 3 prong grounded AC outlet Use other than C B S Scientific s electrical cable with plugs or banana jacks is done...

Page 5: ...utput is controlled by setting a maximum value for voltage up to 300 V or current up to 500 mA The maximum output power at any time is 90 Watts The Run duration can be continuous or programmed to a ma...

Page 6: ...elected parameter 5 RUN Starts and Stops the output from the Power Supply The front and rear panel buttons as shown above are used to set up and run the EPS 300 power supply The audible alarm design a...

Page 7: ...ive Inspect all the components for damage that may have occurred while the unit was in transit If any part appears damaged contact the carrier immediately Be sure to keep all packing material for dama...

Page 8: ...un variable Press the MODE button to check and verify the set value of the other non constant run variable This variable is set by default to the maximum output of the power supply This is the normal...

Page 9: ...clock symbol Press the UP and DOWN buttons to enter in the run time in minutes Press both the UP and DOWN buttons at the same time The display will briefly show PF0 indicating that the PFD mode is dis...

Page 10: ...unction 4 Contact your local C B S Scientific representative Fan is spinning LED Display do not light up Front panel power on indicator off 1 Check main power source 2 Check if any errors on the front...

Page 11: ...etected at some point after a successful start Buffer is leaking from the cells There is a loose electrical connection in the circuit 1 Check and correct all potential resistance problems 2 Check buff...

Page 12: ...de Press any button except the RUN button to clear the error message D 3 Replacing the fuses 1 Unplug the power cord from the power supply 2 Locate the fuse compartment on the rear panel 3 Use a screw...

Page 13: ...pply and allow to dry completely Please contact your local representative or C B S Scientific technical service specialist by E mail technicalservice cbssci com C B S Scientific offers complete techni...

Page 14: ...z 5 200V 01 2A 200W max EPS 300 II Power Supply 120 V 50 60 Hz 10 300V 4 500mA 90W max EPS 300 IIV Power Supply 230 V 50 60 Hz 10 300V 4 500mA 90W max EPS 600 Power Supply 96 240 VAC 50 60 Hz 10 600V...

Page 15: ...C B S Scientific Company P O Box 856 Del Mar CA 92014 Phone 858 755 4959 email technicalservice cbssci com www cbsscientific com Printed in Taiwan...

Reviews: