background image

NH

3

 R

ESPONDER

 

Operating Manual

 

12 

Calibration and Alarm Setpoints 

 
 

Calibration Technologies provides rapid 

turnaround calibration service, which 

includes a certificate of calibration that 

satisfies OSHA PSM documentation 

requirements for the detector.  Six 

month intervals are recommended.   

Call 866-394-5861 to arrange shipment.  

 
 

Guidelines  

When calibrating the detector, adhere to the following 

guidelines: 

 

  The detector is shipped with a factory calibration.  

Calibrate the detector on a regular schedule, 

depending on use and sensor exposure to poisons 

and contaminants.  CTI recommends once every 

180 days (6 months) under normal use. 

  Required gas cal gas: 

PID: 250 ppm Ammonia, balance air 

LEL: 50% LEL Methane (2.5%)  

  Special calibration gas mixes and kits are available 

from Calibration Technologies.  

  If a certified calibration is required, contact 

Calibration Technologies. 

 

 

Calibration Procedure (AutoCal)

 

Verify that the calibration gas being used matches the 

span concentration values that are set for the detector. 

Refer to 

Span Gas Value 

section on page 8. Correction 

factors are not applied during calibration. Correction 

factors that were set prior to calibration are restored 

when the detector returns to normal operation. 

 

Note: Selecting sensor(s) to be calibrated (step #3) is 

not necessary for calibration of the PID and LEL 

sensors as the cal gas being used will only be detected 

by its corresponding sensor. 

 

Start Calibration 

1.  To enter calibration, press and hold   and 

 

simultaneously as the detector beeps, flashes, and 

vibrates to the corresponding countdown. The 

detector then reads

 

Starting calibration

 

Auto Zero 

2. 

Auto Zero

 flashes while the detector automatically 

zeroes the sensors. Do not apply calibration gas 

during this process, otherwise the auto zero step will 

fail.

 

 

 

 

 

 

 

 

 

Summary of Contents for NH3 RESPONDER

Page 1: ...QUICK REFERENCE GUIDE NH3 RESPONDER Portable Ammonia Leak Detector...

Page 2: ...NH3 RESPONDER Operating Manual 2 Specially configured GasAlert Micro 5 PID from...

Page 3: ...Value 8 Correction Factor 9 Datalogger 9 Alarms 10 Viewing Gas Exposures 10 Clearing Gas Exposures 10 Gas Alarm Setpoints 10 Viewing the Alarm Setpoints 10 Stopping a Gas Alarm 11 Pump Alarm 11 Low B...

Page 4: ...mmonia leaks Think of it as an electronic sulfur stick for locating difficult to find leaks by sniffing around valve packing sight glasses shaft seals etc OSHA recommends to bump test the sensor befor...

Page 5: ...ld until the countdown is complete To clear the TWA STEL and MAX gas exposure readings press and simultaneously and hold until the countdown is complete To view the date and time alarm setpoints TWA S...

Page 6: ...tended beep and flash and displays 0 before deactivating User Options Menu If the detector is passcode protected a passcode must be entered to access the user options menu For more information refer t...

Page 7: ...causes the low and high gas alarms audible visual and vibrator to persist until they are acknowledged Press to acknowledge the alarm The detector is shipped with the latch option disabled Passcode Pro...

Page 8: ...he fail alarm The sensor should be replaced and enabled as soon as possible The detector will function normally with the remaining enabled sensors If a sensor is enabled but it is not installed in the...

Page 9: ...selected gas types Datalogger The Logger option is used to define how often the detector records a datalog sample once every 1 to 127 seconds The datalogger records information that can be compiled to...

Page 10: ...rwise the MAX TWA and STEL exposure readings will not clear Gas Alarm Setpoints The gas alarm setpoints trigger the gas alarms and are described in the list below Low alarm Toxics and combustibles Amb...

Page 11: ...again Pump Alarm If the pump stops operating or becomes blocked the detector activates the pump alarm The pump alarm continues until the blockage is cleared or it is acknowledged by pressing If is pr...

Page 12: ...gies If a certified calibration is required contact Calibration Technologies Calibration Procedure AutoCal Verify that the calibration gas being used matches the span concentration values that are set...

Page 13: ...Press to apply new cal gas repeat step 4 Press to end span Remove hose and turn off cal gas The display then advises to press to set or to bypass the calibration due dates Calibration Due Dates Press...

Page 14: ...AA batteries 3 7 Battery shell 1 Open the latch on the bottom of the detector 2 Remove the battery pack by lifting the bottom of the pack away from the detector 3 Unscrew the two captive screws on th...

Page 15: ...s silicon vapors halogenated compounds etc Protect the combustible sensor from exposure to lead compounds silicones and chlorinated hydrocarbons Although certain organic vapors such as leaded gasoline...

Page 16: ...does not activate or automatically deactivates Possible cause Depleted batteries Solution Replace batteries The activation self test fails Possible cause PID Sensor failure Sensor may have exhibited...

Page 17: ...ystal display LCD Backlight Automatically activates whenever there is insufficient light to view the LCD if enabled and during alarm conditions Self test Initiated upon activation Calibration Every 6...

Page 18: ...utable to repair of the product by any person other than an authorized dealer or the installation of unapproved parts on the product or The obligations set forth in this warranty are conditional on a...

Page 19: ......

Page 20: ...NH3 Responder 20170123...

Reviews: