GasAlertMicroClip
Operator’s Manual
14
User Options Menu
To modify the user options, connect the detector to the IR Link
adapter and open Fleet Manager II. Refer to the
Fleet Manager
II Operator’s Manual
for complete instructions.
The following are the available user options:
1.
Sensors
(H
2
S, CO, LEL, and O
2
)
•
Sensor Disabled:
Disables the sensor.
•
Calibration Gas (ppm) / (%LEL) / (%O
2
):
Defines the
calibration gas concentration for each sensor.
•
Calibration Interval (days):
Defines how often a
calibration should be performed.
•
Bump Interval (days):
Defines how often a bump test
should be performed.
•
Low Alarm (ppm) / (%LEL) / (%O
2
):
Defines the low
alarm setpoint.
•
High Alarm (ppm) / (%LEL) / (%O
2
):
Defines the high
alarm setpoint.
•
TWA Alarm (ppm):
Defines the time-weighted average
(TWA) alarm setpoint. H
2
S and CO sensors only.
•
STEL Alarm (ppm):
Defines the short-term exposure limit
(STEL) alarm setpoint. H
2
S and CO sensors only.
•
STEL Interval (minutes):
Defines the short-term
exposure limit (5-15 minutes). H
2
S and CO sensors only.
•
Auto-Zero Startup:
When enabled, the detector
automatically zeros the H
2
S, CO, and LEL sensors during
the startup self-test.
•
LEL By Volume CH4:
When enabled, the LEL reading is
displayed as %vol., assuming a methane environment.
•
O
2
Auto-Calibration on Startup:
When enabled, the O
2
sensor is automatically calibrated during startup.
2.
User Options
•
Confidence Beep:
When enabled, the detector beeps
once every second to verify the battery has sufficient
power to detect hazardous gas and emit an alarm.
Detector is shipped with Confidence Beep disabled.
•
Latching Alarms:
When enabled, the audible, visual, and
vibrator alarms persist during a high or low alarm until the
gas concentration is below the low alarm setpoint and has
been acknowledged by pressing
C
.
•
Safe Mode:
When enabled,
SAFE
continuously displays
on the LCD when all gas concentrations are normal or
below the alarm setpoints.
•
Stealth Mode:
When enabled, the audible alarm, LEDs
and backlight are disabled.
displays on the LCD.
•
Low Alarm Acknowledge:
When enabled, the audible
alarm can be disabled during a low alarm. The vibrator,
LEDs, and LCD remain enabled. For H
2
S, CO, and LEL
sensors only.
•
Datalog Interval (seconds):
Enter a value
(
1-120
seconds).
www.
GlobalTestSupply
.com
Find Quality Products Online at: