Installation | 5
Instruction Manual COBRA NC 0100-0300 B ACV_EN_en
15 | 48
●
Connect the dilution gas connection (DGC) to the gas supply.
Connection size:
– G1/4, ISO 228-1
●
Electrically connect the solenoid valve (MV), see Wiring Diagram Solenoid Valve [
24].
●
Make sure that the gas complies with the following requirements:
Gas type
Dry nitrogen
Gas temperature
°C
0 ... 60
Maximum gas pressure
bar
13
Recommended pressure setting at the pres-
sure regulating valve (PRV)
bar
0.5
Filtration
µm
5
Recommended flow rate
SLM (standard li-
ter per minute)
20
5.2.5
Purge Gas System Connection (Optional)
PRV
PGC
MAN
FME, FR
NIP
MV
INF
Description
PGC
Purge gas connection
FME
Flow meter
FR
Flow regulator
INF
Inlet flange
MAN
Manometer
MV
Solenoid valve
NIP
Nitrogen panel
PRV
Pressure regulating valve
●
Connect the purge gas connection to the gas supply.
Connection size:
– G1/4, ISO 228-1