background image

R

6

en

stopping. Calculate the time the stress should be increased over and practice before testing. The table

shows how many seconds each turn should take - see Table 1. Some Standards allow a slower rate (more

seconds) up to a maximum time.

Table 1: 

WARNING: There will be a loud bang and the Adhesion Tester may jump up a few centimetres

as the dolly and coating detach and the energy in the tester is released. For personal safety it

is recommended that the operator does not lean over the Adhesion Tester during use.

4. Immediately after the test is complete and the pull-off force has been recorded, slacken the handwheel

or nut to remove all the force from the unit.

2.4 ASSESSING THE RESULTS

To record the result, first inspect the face of the dolly. In most cases the coating will fully adhere to the dolly

and the test can be claimed as 100% valid. In some cases, the coating will cover only part of the area of the

dolly and a partial adhesion failure should be recorded.

Elcometer 106

Stress Rate, s/turn (see Standard)

Scale

Max MPa

MPa/turn

1MPa/s

0.2MPa/s

1

3.5

0.21

0.2

1.1

2

7

0.42

0.4

2.1

3

15

1.82

1.8

9.1

4

22

3.75

3.8

18.8

106_TMA_0003_00_09.fm  Page 6  Tuesday, July 31, 2012  2:08 PM

106 

Pull Off Adhesion Tester

8 of 12

For more information

    /    1300 287 935

 

[email protected]    /    www.burwell.com.au

Summary of Contents for Elcometer 106

Page 1: ...Operating Instructions 106_TMA_0003_00_09 fm Page 1 Tuesday July 31 2012 2 08 PM Product Manual 106 Pull Off Adhesion Tester For more information 1300 287 935 mail burwell com au www burwell com au Co...

Page 2: ...ocument may be reproduced transmitted transcribed stored in a retrieval system or otherwise or translated into any language in any form or by any means electronic mechanical magnetic optical manual or...

Page 3: ...7 4 Dollies 8 5 Maintenance 9 6 Technical specification 10 7 Spare parts and accessories 12 8 Related equipment 12 106_TMA_0003_00_09 fm Page 1 Tuesday July 31 2012 2 08 PM 106 Pull Off Adhesion Test...

Page 4: ...ng paint plastic sprayed metal epoxy wood veneers laminates on wood metal or plastic The Adhesion Tester employs a pull off method to determine the force required to pull an area of coating away from...

Page 5: ...Scale 1 0 to 3 5 N mm 500 psi Elcometer 106 2 Scale 2 0 to 7 N mm 1000 psi Elcometer 106 3 Scale 3 0 to 15 N mm 2000 psi Elcometer 106 4 Scale 4 0 to 22 N mm 3200 psi Elcometer 106 5 Scale 5 0 to 0 2...

Page 6: ...essure to squeeze out excess adhesive which should then be removed Allow the adhesive to cure for the recommended time see Adhesives on page 7 Dollies can be re used after cleaning Additional dollies...

Page 7: ...wheel or nut slowly and evenly to apply an increasing force to the dolly and hence stress to the coating Continue until the coating fails and the dolly is removed from the surface or until the specifi...

Page 8: ...mplete and the pull off force has been recorded slacken the handwheel or nut to remove all the force from the unit 2 4 ASSESSING THE RESULTS To record the result first inspect the face of the dolly In...

Page 9: ...ldite which is a two component epoxy paste When mixed it is to be used within 1 hour The two components should be mixed in roughly equal portions Curing Times 24 hours at 25 C 3 hours at 60 C Suitable...

Page 10: ...cure before disposal For further guidance on the disposal of adhesives contact your local environmental authority 4 DOLLIES Dollies can be re used after cleaning Large dollies Coatings on concrete ce...

Page 11: ...the tester is tightened fully beyond the length of the engraved scale on the barrel In this instance the force recorded by the tester should be checked and if necessary the tester should be returned...

Page 12: ...asia elcometer com www elcometer com ENGLAND Elcometer Limited Edge Lane Manchester M43 6BU Tel 44 0 161 371 6000 Fax 44 0 161 371 6010 e mail sales elcometer com www elcometer com USA Elcometer Inc...