E-12-
8205
pH CONTROLLER 8205
3 INSTALLATION
3.1.4 pH controller type 8205 Wall-mounted version installation
The pH controller in wall-mounted version has 4 fixing holes in the bottom enclosure.
Remove the white blanking stripes and the cover to access to fixing holes
.
Fuse
125 mA T
Switch for power
supply selection
230/115 VAC
3.2 Electrical connection
3.2.1 General electrical requirements
The connecting line conducts the measuring signal and must not be installed in combination
with high voltage or high frequency carrying lines. If a combined installation cannot be
avoided, either keep a min. space of 30 cm (approx. 1 ft) or use shielded cables. When
using shielded cables observe faultless grounding of the shield. For normal operating
conditions, the 4-20 mA output, and relays signals can be transmitted by a simple cable
of 0.75 mm
2
. In case of doubt, always use a shielded cable.
The power supply must be of good quality (filtered and regulated).
!
For EMC purposes,the earth must
be connected to the controller.
Fig 3.3 Installation pH controller type 8205 wall-mounted version