Low Battery Alarm and Low Flow Alarm

The EVAHL1 Blower unit is equipped with a Low Battery Alarm and a   
Low Flow Alarm. 
The Low Battery Alarm will sound an intermittent electronic beep indicating that 
there are approximately 15 minutes of remaining battery capacity.  The delays 
between beeps will get shorter and shorter as time runs out.
The Low Flow Alarm will sound a continuous electronic beep indicating that the 
flow to the hood has dropped below the design specification of 185 lpm = 6.5 CFM 
(Note: The NIOSH minimum required flow is 170 lpm = 6 CFM). 
When either of these alarms sounds, the user should immediately: 
  •Leave the hazard area
  •Remove the headpiece
  •Disconnect the breathing tube from the hood 
  •Check the airflow with the airflow indicator (see page 4). 
  • Check the operation of the low-flow alarm by blocking the end of the 

breathing tube. The device will first ramp up to compensate and if correct 
flow cannot be achieved, the alarm will sound within 5 seconds.

If the airflow indicator indicates insufficient airflow, the battery should be fully 
charged (see “Battery Pack” section), and/or the filter/cartridge should be 
replaced. 

  

NOTE 

  

 The EVAHL1 blower is provided with a circuit to protect the battery. It 
will not allow the battery to be discharged below a safe voltage for the 
cells, regardless of airflow, without the Alarm sounding. When the battery 
reaches the voltage cutoff it will automatically cease operation.   

Doffing the Respirator

Prepare to doff the blower, battery and hood in a safe, hazard-free area and 
do the following (in conjunction with your employer’s standard operating 
procedures): 
• Remove the hood.
•  Turn the blower off by holding down the ON/OFF switch for 5 seconds.  This is 

confirmed by a long beep and a shut down of the motor.

• Remove the waist belt. 
• Disconnect the hood from the breathing tube.
• Disconnect the breathing tube from the blower.
• Clean and inspect components as necessary.

Troubleshooting 

The following guide will assist you in troubleshooting to locate possible issues with your respirator: 

Circumstance

Possible Cause(s)

Solution

Low Battery Alarm is sounding

Low Voltage

Charge the battery

Blower malfunction

Return blower for anyalysis

Low Flow Alarm is sounding

Clogged/damaged air-purifying filter element

Replace the filter/cartridge

Battery Low

Re-charge the battery

Blower malfunction

Leave hazardous area immediately and check 
equipment. If the problem persists and no damage 
is found, return equipment for repair. 
Replace breathing tube and/or hood.

Hood neck cuff is restricting flow

Adjust neck cuff position

Smell or taste contaminant

Equipment damaged

Leave hazardous area immediately and check 
equipment

Filter needs to be replaced

Replace filter

Low airflow

Leave hazardous area immediately and check 
equipment
If the problem persists and no damage is found, 
return equipment for repair

Blower unit does not run full service life

Damaged Battery

Return battery for analysis

Malfunctioning Battery Charger

Return charger for analysis

Hood neck cuff is restricting flow

Adjust neck cuff position

Summary of Contents for EVAHL Series

Page 1: ...t use these respirators for respiratory protection during abrasive blasting or clean up 4 Do not use in circumstances where the airborne concentration level of contaminant exceeds maximum use concentration for this type of respirator as established by regulatory standards 5 Leave area immediately if Breathing becomes difficult Dizziness or other distress occurs You taste or smell the contaminant U...

Page 2: ...edium size 20LF2L Loose fitting facepiece narrow profile large size 6 The Battery Charger EVAHLSMC Quick charger single port EVAHLGC Gang charger six port The blower unit draws in ambient air through the cartridges The purified air is blown into the wearer s hood through the breathing tube A flow indicator is provided to check that there is an adequate volume of air available to the wearer prior t...

Page 3: ...terioration physical damage and improper assembly 2 Filter Cartridges Inspect the filter cartridge for any physical damage Check the label to ensure the filter cartridge has not exceeded its use by date Inspect the gasket on the filter for any phsyical damage NOTE Each filter comes with a permanent gasket Ensure that the correct filter cartridge is appropriate for the contaminant Consult the NIOSH...

Page 4: ...particulate filters are 99 97 effective against all particulate aerosols The following abbreviations are approved by NIOSH to indicate the particulates gases or vapors which are removed by the gas vapor cartridges HE High Efficiency Particulate OV Organic Vapor AM Ammonia CD Chlorine Dioxide CL Chlorine FM Formaldehyde HC Hydrogen Chloride HF Hydrogen fluoride MA Methylamine SD Sulfur Dioxide Moun...

Page 5: ...ead during use first verify proper air pressure then select a different hood for your application or use the optional chin strap Adjusting and Installing Hard Hat in Respirator Hood 20SICH 20TICH 1 Assemble and adjust the standard Bullard hard hat suspensions RS4PC or RS6PC or the optional ratchet suspensions RS4RC or RS6RC by following the directions on instruction sheet attached to headband on h...

Page 6: ...hing 6 Ensure that the neck cuff is down below the chin and that the air outlets of the cuff see Figure 27 are not restricted If the neck cuff is not below the chin then pull down before continuing See Figure 27 RT Series Hood Use Installing Breathing Tube Assembly in RT Series Respirator Hoods For hoods without a threaded port at the rear Breathing Tubes PA1BT PA1BTXS and PA1BTXL will attach to t...

Page 7: ...ttach the other end of breathing tube to blower unit if not already attached by screwing adapters together Remove any protective film covering the lens of the headpiece Put on the belt and blower assembly and make any final adjustments to the belt as necessary keeping the breathing tube and hood behind the head Turn the blower on by depressing and holding the on off switch Figure 33 for approximat...

Page 8: ...will automatically cease operation Doffing the Respirator Prepare to doff the blower battery and hood in a safe hazard free area and do the following in conjunction with your employer s standard operating procedures Remove the hood Turn the blower off by holding down the ON OFF switch for 5 seconds This is confirmed by a long beep and a shut down of the motor Remove the waist belt Disconnect the h...

Page 9: ...SSEMBLIES ALTERNATE SUSPENSION HARD HAT ASSEMBLIES ALTERNATE BREATHING TUBES ALTERNATE BLOWER UNIT ASSEMBLIES FILTER ASSEMBLY ALTERNATE CARTRIDGE ASSEMBLIES ALTERNATE BATTERY CONFIGURATIONS ALTERNATE BELT ASSEMBLIES ALTERNATE CHARGER ASSEMBLIES ACCESSORIES CAUTIONS AND LIMITATIONS2 20LFXL 20LFL 20LFM 20LFXLHE 20LFLHE 20LFMHE 20LF2L 20LF2M 20LF2S 20LF2LHE 20LF2MHE 20LF2SHE 20TJN 20TJNT 20TICN 20TIC...

Page 10: ...ullard will accept any returned goods Please read carefully Follow the steps outlined below to return goods to Bullard for repair or replacement under warranty or for paid repairs 1 Contact Bullard Sales Support by telephone or in writing at Bullard 1898 Safety Way Cynthiana KY 41031 9303 Toll free 877 BULLARD 285 5273 Phone 859 234 6616 In your correspondence or conversation with Sales Support de...

Page 11: ...cepieceTychem 2000Traditional Large HEPA 20LFMHE 20LF LF FacepieceTychem 2000Traditional Medium HEPA 20LFXLHE 20LF LF FacepieceTychem 2000Traditional XL HEPA Double bib hood for use without a headband suspension RT1 RT1T Tychem 2000 QC Inflatable neck collar RT2 RT2T Tychem 4000 SL Inflatable neck collar RT3 RT3T Tychem 2000 QC Sport neck collar RT4 RT4T Tychem 4000 SL Sport neck collar Accessory ...

Page 12: ...59 234 6616 Fax 1 859 246 0243 Americas Operations 1898 Safety Way Cynthiana KY 41031 USA 877 BULLARD 285 5273 Tel 1 859 234 6616 Fax 1 859 234 8987 Bullard GmbH Dieselstrasse 8a 53424 Remagen Germany Tel 49 2642 999980 Fax 49 2642 9999829 Bullard Asia Pacific Pte Ltd 51 Changi Business Park Central 2 03 04The Signature Singapore 486066 Tel 65 6745 0556 Fax 65 6797 0299 ...

Reviews: