TC-MIDI (+)
9 Appendices
9.1 Gas cooler technical data
TC-MIDI
Gas Cooler Technical Data
Ready for operation
after max. 10 minutes
Ambient temperature
5 °C to 60 °C
Gas outlet dew temperature
preset:
adjustable:
5 °C
2 °C…20 °C or Delta T-control
Protection class
IP 20
Mechanical load
Tested to DNVGL-CG-0339, Table 6
1)
2 Hz-13.2 Hz Amplitude ± 1.0 mm
13.2 Hz -100 Hz acceleration
Housing
Stainless steel, brushed
Packaging dimensions
approx. 350 x 220 x 220 mm
Weight incl. heat exchanger
approx. 11,5 kg
approx. 15 kg at full expansion stage
Electrical data
Unit without add-on
Unit with add-on
(P1.x + Peristaltic pump)
230 V AC
115 V AC
230 V AC
115 V AC
1.2 A
2.4 A
1.8 A
3.6 A
200 W / 280 VA
290 W / 420 VA
Recommended fuse (characteristic: delayed action)
3,15 A
6,3 A
3,15 A
6,3 A
Status output switching capacity
max. 250 V AC, 150 V DC
2 A, 50 VA, potential-free
Electrical connections
Plug per EN 175301-803
Gas connections and condensate outlet
Heat exchanger see table "Heat Exchanger Overview"
Filter, moisture detector adapter G1/4 or NPT 1/4"
Parts in contact with mediums
Filter:
Moisture detector:
Heat exchanger:
Peristaltic pump:
Sample gas pump:
Tubing:
see "Technical Data - Options"
see "Technical Data - Options"
see table "Heat Exchanger Overview"
see "Technical Data - Options"
see "Technical Data - Options"
PTFE/Viton
FM No.
3062014
1)
not in conjunction with add-on sample gas pump
34
Bühler Technologies GmbH
BE440025 ◦ 10/2018