![Buhler TC-MIDI 6111 Installation And Operation Instructions Manual Download Page 43](http://html1.mh-extra.com/html/buhler/tc-midi-6111/tc-midi-6111_installation-and-operation-instructions-manual_2827967043.webp)
TC-MIDI (+)
TC-MIDI+
The energy content of the sample gas and the required cooling capacity of the gas cooler is determined by three parameters: gas
temperature
ϑ
G
, dew point
τ
e
(moisture content) and volume flow v. The outlet dew point rises with increasing energy content of
the gas. The approved energy load from the gas is therefore determined by the tolerated rise in the dew point.
The following limits are specified for a standard operating point of
τ
e
= 50 °C and
ϑ
G
= 70 °C. The maximum volume flow v
max
in
Nl/h of cooled air is indicated, so after moisture has condensed.
If the values fall below
τ
e
and
ϑ
G
, the flow v
max
may be increased. For example, on the STG-2 heat exchanger the parameter triple
τ
e
= 40 °C,
ϑ
G
= 70 °C and v = 575 Nl/h may also be used in place of
τ
e
= 50 °C,
ϑ
G
= 70 °C and v = 320 Nl/h.
Please contact our experts for clarification or refer to our design program.
9.6.2 Heat exchanger overview
TC-MIDI
Heat exchanger
TS
TS-I
2)
TG
TG
TV-SS
TV-SS-I
2)
DTS (DTS-6
3)
)
DTS-I (DTS-6-I
3)
)
2)
DTG
DTG
DTV
3)
DTV-I
2) 3)
Version / Material
Stainless
steel
Glass
PVDF
Stainless steel
Glass
PVDF
Flow rate v
max
1)
500 L/h
400 L/h
235 L/h
2 x 250 L/h
2 x 200 L/h
2 x 160 L/h
Inlet dew point
τ
e,max
1)
80 °C
80 °C
65 °C
80 °C
65 °C
65 °C
Gas inlet temperature
ϑ
G,max
1)
180 °C
140 °C
140 °C
180 °C
140 °C
140 °C
Max. Cooling capacity Q
max
450 kJ/h
230 kJ/h
120 kJ/h
450 kJ/h
230 kJ/h
185 kJ/h
Gas pressure p
max
160 bar
3 bar
3 bar
25 bar
3 bar
2 bar
Pressure drop Δp (v=150 L/h)
8 mbar
8 mbar
8 mbar
5 mbar each
5 mbar each
15 mbar each
Dead volume V
tot
69 ml
48 ml
129 ml
28 / 25 ml
28 / 25 ml
21 / 21 ml
Gas connections (metric)
G1/4
GL 14 (6 mm)
4)
DN 4/6
6 mm tube
GL14 (6 mm)
4)
DN 4/6
Gas connections (US)
NPT 1/4"
GL 14 (1/4")
4)
1/4"-1/6"
1/4" tube
GL14 (1/4")
4)
1/4"-1/6"
Condensate out connections (metric)
G3/8
GL 25 (12 mm)
4)
G3/8
Tube 10 mm
(6 mm)
GL18 (10 mm)
4)
DN 5/8
Condensate out connections (US)
NPT 3/8"
GL 25 (1/2")
4)
NPT 3/8"
Tube 3/8" (1/4")
GL18 (3/8")
4)
3/16"-5/16"
1)
Max. cooling capacity of the cooler must be considered
2)
Models marked I have NPT threads or US tubes, respectively.
3)
Condensate drain only possible with condensate pump
4)
Gasket inside diameter
TC-MIDI+
Heat exchanger
2x STG-2
2x STV-2
Version/Material
Glass
PVDF
Flow v
max
1)
320 L/h
300 L/h
Inlet dew point
τ
e,max
1)
70 °C
70 °C
Gas inlet temperature
ϑ
G,max
1)
140 °C
140 °C
Max. Cooling capacity Q
max
345 kJ/h
210 kJ/h
Dead volume V
tot
47 ml
41 ml
Gas connections (metric)
GL 14 (6 mm)
2)
DN 4/6
Gas connections (US)
GL 14 (1/4")
2)
1/4”-1/6“
Condensate out connection (metric)
GL 18 (10 mm)
2)
G1/4
Condensate out connection (US)
GL 18 (10 mm)
2)
NPT 1/4“
1)
Max. cooling capacity of the cooler must be considered
2)
Gasket inside diameter
41
Bühler Technologies GmbH
BE440025 ◦ 10/2018