TC-Double (+) X2
5.3 Description of menu functions
5.3.1 Main menu
Peltier cooler TC-Double (tc.do)
Display →
tc.do
This will take you to the cooler target temperature and the tolerance range setting (alarm
threshold).
Global setting (ToP Settings)
Display →
toP
This menu is used to configure the global cooler settings.
Peristaltic Pump
Display →
PuMP
Switching the peristaltic pump on and off.
Parameter range:
NoP
,
oFF
Factory setting:
NoP
Note:
Status switches, "
PuMP
" flashes.
Heat exchanger material selection
Display →
h.chn
Heat exchanger material selection
Parameter range:
tS
(Steel),
t6
(Glass),
tU
(PVDF)
Factory setting:
tS
(cooler without heat exchanger), or respective material per configuration
Exit main menu
Display →
E
Selecting this will return you to display mode.
5.3.2 Submenu 1
Target temperature (Temperature)
Display → Cooler →
temp
This setting determines the nominal temperature for the cooler temperature.
Parameter range:
2 °C to 20 °C (35.6 °F to 68 °F)
Factory setting:
5 °C (41 °F)
Note:
If the temperature is changed the indicator may blink, until the new operating range has been
reached.
This menu item is hidden if the keylock is enabled.
20
Bühler Technologies GmbH
BE440022 ◦ 10/2018