4.2.1 Overview of the Menu Items
Display of current temperature and operating state
Peltier Gas Cooler
Display:
P5__
2°C...20°C
Display:
xxxx
Global Settings
Display:
toP
+1
°C…+7C
Display:
xxx
Exit Main Menu
Display:
E
t b
Exit Submenu
Display:
E
j
Main Menu
Set Temperature
Display:
tEMP
Submenu
upper alarm threshold
Display:
A Hi
Setting Range
Celsius/ Fahrenheit
Display:
C
[F]
j
j
t b
t b
t b
j
j
j
j
Current Temperature
Display:
xxxx
Displayed is the block temperature with a resolution of 0.5°C. By pressing the Enter button brings the display to
the Main Menu. The unit of temperature is adjustable in the menu Global settings (Celsius or Fahrenheit).
Adjust the set point:
2°C...20°C
Set the upper alarm
Threshold above setpoint:
tEMP +1°C...+7°C
Set unit to Celsius or
Fahrenheit.
-1
°C…-3C
Display:
xxx
lower alarm threshold
Display:
A Lo
j
t b
Set the lower alarm
Threshold below setpoint:
tEMP -1°C...-3°C
t b
adjust value
j
store value
wait 4s: no storage
VE 00 0001
Fig. 1: Overview of the menu