Buckleys BathyCorrometer Pro' Installation Manual Download Page 3

1. 8mm open-ended spanner

dealer to source

2. 19mm open-ended spanner

dealer to source

3. Thermometer (K-type)/salinity meter

6005-0996

4. 2m/kG calibrated torque wrench 

6005-0997

with 19mm socket (½” square drive)

5. Digital volt meter and leads

dealer to source

6. BathyCorrometer

®

service tool (T-bar) 

6005-0313

7. BathyCorrometer

®

locking ring tool

6004-0007

8. Pipe grips

dealer to source

9. De-ionized water (10L)

dealer to source

10. K-Cell adaptor leads

6004-0018

11. Tamper-evident seal/paint

dealer to source

12. Crimp tool

JST WC-240 (Farnell 3617270)

13. BathyCorrometer

®

Calibration Kit

6004-0031

14. Soldering Iron

dealer to source 

15. Wire cutters 

dealer to source 

16. Long nose pliers 

dealer to source 

17. Lead-free solder 

dealer to source 

18. Cleaning cloth 

dealer to source 

19. Hammer

dealer to source 

20. Handle sealing plug 

6005-0312

21. Half-cell sealing plug

6005-0311

22. Trim Tool

6005-0084

23. Suction puller cup

2006-0036

24. Pressure tank capable of 600psi

dealer to source

2.

Preparation

Workspace

A robust workspace/bench is required and should be clean and free from debris. All testing and calibration equipment such as pressure chambers 
and multimeters should be checked and calibrated annually by recognised or approved sources.

Tools required

In order to check and fully test a BathyCorrometer

®

Pro’, the following service and calibration tools are required:

Items with part numbers may be 
purchased from Buckleys

1

2

3

4

5

6

7

8

16

17

14

15

12

13

9

11

23

24

21

22

19

20

10

18

BathyCorrometer

®

M k V

t o   P R O ’   U P G R A D E  M A N U A L

HOME

Reassembly

Description

Preparation

Disassembly

Repairs & test certificate production

Summary of Contents for BathyCorrometer Pro'

Page 1: ...o Upgrade Kit Installation manual Or press Esc on your keyboard to close this manual 1 Description 4 Reassembly 2 Preparation 3 Disassembly Click on a title below to go to that section 5 Repairs test...

Page 2: ...MkV units up to that of the current model with improved display clarity the correct negative potential reading and revised battery and electronics systems BathyCorrometer MkV to PRO UPGRADE MANUAL HOM...

Page 3: ...e solder dealer to source 18 Cleaning cloth dealer to source 19 Hammer dealer to source 20 Handle sealing plug 6005 0312 21 Half cell sealing plug 6005 0311 22 Trim Tool 6005 0084 23 Suction puller cu...

Page 4: ...g ring Using a suitable spanner or wrench and socket unscrew and remove the locking ring by turning it counter clockwise Some assistance to hold the BathyCorrometer is advised whilst the locking ring...

Page 5: ...ter on a firm surface and locate the locking ring removal tool s three locating pins in to the holes on the locking ring Carefully retrieve the lens locating pins part no 3009 0010 and remove the larg...

Page 6: ...ronics assembly from the BathyCorrometer to gain sufficient access to disconnect the plug connection on the rear of the assembly Withdraw the complete electronics unit from the BathyCorrometer Remove...

Page 7: ...ve the white multi plug by cutting the wires as close to the plug as possible BathyCorrometer MkV to PRO UPGRADE MANUAL Cut wires as close to multiplug as possible HOME Reassembly Description Preparat...

Page 8: ...orrect order See the illustration for the correct multi plug wiring order IMPORTANT incorrect wiring of the 6 way multiplug is likely to cause irreparable damage to the BathyCorrometer Pro s electroni...

Page 9: ...e electronics assembly so the top and bottom holes align with the fixing points in the body of the BathyCorrometer Insert and tighten the two self tapping screws part no 2003 0174 until the electronic...

Page 10: ...6nm using the locking ring tool and a calibrated torque wrench with an appropriate socket Pressure test Pressure test at 600psi for 15 minutes After pressure testing thoroughly dry the outside of the...

Page 11: ...completion of the installation the original battery charger and BCM Checker must be replaced with the BathyCorrometer Pro Charger and the CalChecker Pro Replacement parts For details of all replacemen...

Page 12: ...BatheyCorrometer Pro 2 Click on My Customers 1 Go to www buckleysinternational com and click on SIGN IN button to log in to your Buckleys account with your username and password BathyCorrometer MkV t...

Page 13: ...stomer complete the registration form and then click Add Customer Return to My Customers page 4 Enter the customer s details and click Add Customer at the bottom of the page BathyCorrometer MkV to PRO...

Page 14: ...he field below What is the product s serial number and enter the product s unique serial number Enter details for the work and procedures undertaken and parts used and click Continue It is important t...

Page 15: ...our records and you will now have the option to print a copy of the certificate or download a PDF Please print a copy of the certificate and include it with the product for return to the customer Repa...

Reviews: