4.
Reassembly
BathyCorrometer
®
M k V
t o P R O ’ U P G R A D E M A N U A L
Verification test
Place the BathyCorrometer
®
in to a solution of 3% salinity de-ionised water at 25°C, keeping the bulkhead socket above the waterline and dry.
Connect a digital Voltmeter to the socket and take readings using 3 laboratory K-series reference cells
The new electronics should give a negative reading of 42mV (+/-5mV) and a zinc block test should give a negative reading of 1.040 (+/-40mV).
Battery charger and CalChecker
Pro’
Upon completion of the installation, the original battery charger and BCM Checker
must be replaced with the BathyCorrometer
®
Pro’ Charger and the
CalChecker
Pro’.
Replacement parts
For details of all replacement parts available, please refer to the BathyCorrometer
®
Pro’ Parts List within the
BathyCorrometer Pro’ Service, Diagnostics and Repair Manual
.
More information
For more information regarding any of the procedures described in this manual, please refer to the
BathyCorrometer Pro’ Service, Diagnostics and Repair Manual
.
IMPORTANT:
Using the old charger or BCM Checker can damage the new electronics.